background image

1

Chapter 1

Introduction

Power Amplifier Type 2719 has been designed to drive small vibration exciters, particularly the
112 N (25 lbf) Vibration Exciter Type 4808. The RMS output-current limit is adjustable, and
consequently Type 2719 can drive the 45 N (10 lbf) Vibration Exciter Type 4809 safely to full
rating. The power amplifier has a usable frequency range from DC to 100 kHz. The rated AC
output is 180 VA into a 0.8

Ω

 exciter or resistive load, in the frequency range DC to 15 kHz

(

±

0.5 dB). The maximum voltage gain is 40 dB. Harmonic content of the output is very small as

heavy negative feedback is used and the instrument can tolerate temperature and supply line
variations while maintaining excellent stability. 

Type 2719 can be used as a voltage generator with low output impedance and a flat voltage
frequency response, or as a current generator with high output impedance and a flat current
frequency response. 

Summary of Contents for 2719

Page 1: ...5 83 05 Ireland 353 1 807 4083 Italy 39 0257 68061 Japan 81 3 5715 1612 Netherlands 31 318 55 9290 Norway 47 66 77 11 55 Poland 48 22 816 75 56 Portugal 351 21 41 69 040 Republic of Korea 82 2 3473 06...

Page 2: ...BE 1660 12 May 2006 Power Amplifier Type 2719 User Manual...

Page 3: ...plosion Hazard The equipment is not designed to be used in potentially explosive environments It should not be operated in the presence of flammable liquids or gases Warnings Whenever it is likely tha...

Page 4: ...oduction 1 CHAPTER 2 Controls 3 Front Panel 3 Back Panel 4 CHAPTER 3 Operation 5 Preliminary 5 System Checks and Connections 5 CHAPTER 4 Characteristics 11 Signal Inputs 11 CHAPTER 5 Specifications 17...

Page 5: ......

Page 6: ...e from DC to 100 kHz The rated AC output is 180 VA into a 0 8 exciter or resistive load in the frequency range DC to 15 kHz 0 5 dB The maximum voltage gain is 40 dB Harmonic content of the output is v...

Page 7: ...Power Amplifier Type 2719 User Manual 2...

Page 8: ...is switched On The Voltage or Current Mode LED indicator lights green showing the output mode The Interlock LED indicator lights red when the interlock circuit is activated The 0 or 180 LED indicator...

Page 9: ...impedance is 10 k Full output is produced by a 2 4 V RMS input signal Voltage Monitor BNC socket providing output of the amplifier voltage waveform including DC component for display on an oscillosco...

Page 10: ...is designed to fit into 19 standard rack system It may be used free standing on its four rubber feet however when fitted in a instrumentation rack the rubber feet must be removed System Checks and Con...

Page 11: ...e amplifier unit is aligned with the appropriate voltage printed on the selector Fuse Check and Replacement The mains fuse and fuse holder are integrated in the voltage selector It can be released usi...

Page 12: ...ther This is done automatically through the screens of the input and output cables used to interconnect the instruments 2 If instruments equipped with a mains socket protective earth terminal are empl...

Page 13: ...exciter is exceeded causing the vibration table to knock against its end stops reduce the generator output voltage to a lower level in order to resume operation For frequency sweep vibration testing...

Page 14: ...Reset and make the necessary adjustments Normal operation can then be resumed by returning Variable Gain dB to the position used for test If a shut down is caused by an internal fault in the amplifier...

Page 15: ...Power Amplifier Type 2719 User Manual 10...

Page 16: ...s capacitive coupled giving a 0 5 dB lower limiting frequency of approximately 15 Hz The DC signal input is direct coupled enabling a DC offset voltage to be applied for centring vibration exciter tab...

Page 17: ...e exciter applications as well as to multiple exciter applications at low frequencies where it is important that the exciters have the same motion In Current Mode high impedance feedback is proportion...

Page 18: ...btained using exciters with a nominal load impedance of 0 8 For other exciter load impedances the maximum output rating is as shown in Fig 4 3 This is valid at frequencies from 5 Hz to 15 kHz At other...

Page 19: ...pedance valid at frequencies from 40 Hz to 10 kHz 200 Maximum Output Power VA Vibration Exciter Type 4808 Vibration Exciter Type 4809 Current Power Voltage Power Current Voltage 180 160 140 120 100 80...

Page 20: ...nput socket used and the setting of the V C Mode switch as shown by the response curves for small signals given in Fig 4 4 Fig 4 4 Frequency response in Voltage Mode of Type 2719 for power output leve...

Page 21: ...s can be attributed to the generous amount of feedback applied and the use of a direct coupled input Fig 4 5 Typical harmonic distortion percentage curves for Voltage Mode and Current Mode output impe...

Page 22: ...TPUT VOLTAGE CAPACITY 12 V RMS DC to 15 kHz via 4 pin Neutrik Speakon plug OUTPUT CURRENT CAPACITY 7 5 A RMS at or below 5 Hz 15 A RMS 40 Hz to 10 kHz 12 A RMS at 15 kHz FREQUENCY RANGE Full Capacity...

Page 23: ...rrent true RMS adjustable limit 1 to 15 A RMS Power Transistor Temperature Heat Sink Temperature Output Signal Distortion no shut down OTHER FEATURES Electronic peak current limiting POWER REQUIREMENT...

Page 24: ...it against unintended operation For repair contact your local Br el Kj r representative Under no circumstances should repair be at tempted by persons not qualified in the service of electronic instrum...

Page 25: ...Power Amplifier Type 2719 User Manual 20...

Page 26: ...5 83 05 Ireland 353 1 807 4083 Italy 39 0257 68061 Japan 81 3 5715 1612 Netherlands 31 318 55 9290 Norway 47 66 77 11 55 Poland 48 22 816 75 56 Portugal 351 21 41 69 040 Republic of Korea 82 2 3473 06...

Reviews: