CHAPTER 4
The Use of Control Software BZ 5291
23
Input Floating
Input Float:
Used to select single-ended or floating (differential) input mode. Select
Yes for floating input.
When DeltaTron mode is selected for DeltaTron/voltage input modules, single-ended
mode must be selected if an alternative return path for the supply current has not
been established.
Software Version
Shows the version number for the boot and application software that constitutes the
current Type 2694 program. These numbers are only changed if the Type 2694 ampli-
fier is updated with completely new versions of these programs.
Software Revision
The revision number shows the revision status for the current boot and application
program versions. These numbers are changed for small updates of the program,
e.g., error corrections, minor new functions and special customer versions.