background image

XT Freezing Core

Working 

Temperature Range

CoolBox XT 

Cooling Source

Freezer 

Temperature

Temperature Duration

*

Open Lid

Temperature Duration

*

Closed Lid

-20° to 0

o

C

XT Freezing Core

-20

o

C

Over 5 hours

Over 8 hours

-20° to 0

o

C

XT Freezing Core

-80

o

C

Over 8 hours

Over 12 hours

Maintaining samples at -20° to 0°C 
Freeze XT Freezing Core in a -20°C freezer for at least 12 hours, or for a faster start, freeze in a -80°C 

XT Cooling Core

Working 

Temperature Range

Temperature Duration

Open Lid

Temperature Duration

Closed Lid

0.5° to 4.0

o

C

Over 10 hours

Over 16 hours

Maintaining samples at 0.5° to 4°C 
Freeze XT Cooling Core in a -20°C freezer for at least 12 hours. XT Cooling Core should be stored 
in -20°C freezer when not in use so it is ready when needed. Note: Freezing XT Cooling Core for less 
than the specified time will result in decreased cooling duration. 

When using a room temperature CoolRack

®

 or CoolSink

®

 modules

•  Remove two XT Cooling Cores from the freezer and place into base.
•  Fit magnetized collar onto base, seating it securely.
•  Place CoolRack

®

 or CoolSink

®

 modules directly onto core and allow to equilibrate to 4°C   

(approximately 10 -15 minutes). 

•  Load samples.
•  Place lid on CoolBox 2XT when not processing samples to maximize cooling duration.

When using charged XT Cooling Cores and pre-chilled (1°C or less) CoolRack

®

 or CoolSink

®

 modules

•  Remove XT Cooling Cores from the freezer and place on benchtop for approximately 10 minutes. 

When the temperature indicator displays 1°C, XT Cooling Core is ready to use. 

IMPORTANT: Failure to allow XT Cooling Cores to reach 1°C may result in undesired sample freezing.
•  Place XT Cooling Cores into base.
•  Fit magnetized collar onto base, seating it securely.
•  Place the CoolRack or CoolSink modules of choice onto the core. No equilibration time is needed.
•  Load samples.
•  Place lid on CoolBox 2XT when not processing samples to maximize cooling duration.

Assembly of CoolBox

 2XT cont. 

1. Remove XT Cooling Cores from the freezer and place 

on benchtop for approximately 10-15 minutes. When 
the temperature indicator displays 1°C, XT Cooling 
Core is ready to use. Note: Frost will form on the core 
exterior upon removal from the freezer; when the frost 
liquifies, the core is at proper temperature.

2. Place two XT Cooling Cores into CoolBox 2XT base.
3. Fit magnetized collar onto CoolBox 2XT base, seating 

it securely.

4. Place two CoolRack

®

 or CoolSink

®

 modules of choice 

onto XT Cooling Cores.

5. Load samples.
6. Place lid on CoolBox 2XT when not processing samples 

to maximize cooling duration.

Lid

Magnetized collar

XT Cooling Core(s) 
or XT Freezing 
Core(s)(Pre-frozen in 
-20˚C freezer
)

Magnetized base

freezer for at least 6 hours. Freezing in a -80°C freezer also prolongs cooling duration.  XT Freezing 
Core should be stored in a -20°C or -80°C freezer when not in use so it is ready when needed. Note: 
Freezing XT Freezing Core for less than the specified time will result in decreased cooling duration.  

When using a room temperature CoolRack

®

 or CoolSink

®

 module

•  Remove XT Freezing Core from the freezer and place into base.
•  Fit magnetized collar onto the base, seating it securely.
•  Place the room temperature CoolRack or CoolSink module of choice onto core and allow to 

equilibrate to 0°C (approximately 10 minutes).  

•  Load samples.
•  Place lid on CoolBox 2XT when not processing samples to maximize cooling duration. 

When using a pre-chilled (1°C or less) CoolRack

®

 or CoolSink

®

 module

•  Remove XT Freezing Core from the freezer and place into base.
•  Fit magnetized collar onto the base, seating it securely.
•  Place the CoolRack or CoolSink module of choice onto core.  
•  Load samples.
•  Place lid on CoolBox 2XT when not processing samples to maximize cooling duration.

Dry Ice 

Working 

Temperature Range

CoolBox XT 

Cooling Source

Temperature Duration

*

Open Lid

Temperature Duration

*

Closed Lid

-78

o

C

200ml dry ice

Over 4 hours

Over 5 hours

Maintaining or snap-freezing samples at -78°C
•  Remove XT Cooling or Freezing Cores from the CoolBox

 2XT base.

•  Fill the base, with approximately 200ml of pulverized dry ice.
•  Fit magnetized collar onto the base, seating it securely.
•  Place the CoolRack or CoolSink modules directly onto dry ice and allow module to equilibrate to 

-78°C (approximately 7-8 minutes).

•  Load samples.
•  If snap-freezing, freezing will occur in 7-10 minutes depending upon sample volume and type.

Note: The thermo-conductive design of CoolRack

®

 and CoolSink

®

 modules ensures uniform well-to- well 

temperature regardless of the consistency of the dry ice.

Care and Cleaning
The CoolBox

 2XT housing is constructed from a cross-linked, closed-cell polyethylene foam. The 

material has excellent resistance to fluid absorption and abrasion. Do not use the CoolBox XT base 
for pulverizing dry ice. Maximum temperature exposure: 60°C. Avoid prolonged exposure to UV 
light sources. 

All components including housing, XT Cooling Core and optional XT Freezing Core are compatible 
with repeated and prolonged cryogenic temperature exposure. All components can be cleaned 
with aqueous detergents, alcohol, 10% bleach, and acid/base viricide (such as Virkon S) solutions. 
Rinse with clear water after using cleaning solutions. Do not autoclave. CoolRack

®

 and CoolSink

®

 

sample modules may be autoclaved, or cleaned with alcohol or 10% bleach.

Dimensions (L x W x H)
Interior (with XT Cooling Core in base): 21.0 x 14.3 x 13.4 cm / 8.3 x 5.6 x 5.3  in 
Exterior: 26.5 x 20.0 x 15.0 cm / 10.4 x 7.9 x 6.3 in

Reviews: