Brookfield Engineering Laboratories, Inc.
Page 5
Manual No. M/92-021-N0902
I.3 Specifications
Speeds:
LVDV-I
+
:
0.0, 0.3, 0.6, 1.5, 3, 6, 12, 30, 60, 0.0,
(rpm)
0.5, 1, 2, 2.5, 4, 5, 10, 20, 50, 100
RV/HA/HBDV-I
+
:
0.0, 0.5, 1, 2, 2.5, 4, 5, 10, 20, 50, 100,
0.0, 0.3, 0.6, 1.5, 3, 6, 12, 30, 60
Weight:
Gross Weight
20
lb
9 kg
Net Weight
17
lb
7.7 kg
Carton Volume
1.65 cu ft
0.05 m
3
Tp
Temperature Sensing Range: -100
°
C to +300
°
C (-148
°
F to +572
°
F)
Operating Environment:
0
°
C to 40
°
C Temperature Range (32
°
F to 104
°
F)
20% - 80% R.H.: non-condensing atmosphere
Analog Torque Output:
0 - 1 Volt DC (0 - 100% Torque)
Tp
Analog Temperature Output: 0-4 Volt DC (10 mV/
°
C)
Viscosity Accuracy:
±
1% Full Scale Range in Use (See Appendix E for details)
Viscosity Repeatability:
0.2% of Full Scale Range in Use
Tp
Temperature Accuracy:
±
1
°
C: -100
°
C to +149
°
C
±
2
°
C: +150
°
C to +300
°
C
Electrical Certifications:
Conforms to CE Standards:
BSEN 50081-1: Emission Standard - Light Industrial
BSEN 50082-1: Immunity Standard - Light Industrial
BSEN 50081-2: Emission Standard - Industrial
BSEN 50082-2: Immunity Standard - Industrial
BSEN 61010-1: Safety requirements for electrical
equipment, for measurement, control
and laboratory use
Approved Standards:
CSA Std. C22.2 No. 151-M1986 - Laboratory Equipment
CSA Class 8721 01 - Laboratory Equipment
Installation Category (over-voltage category) II: Classification of parts of installation
systems or circuits in local level, portable equipment, appliances, etc.
I.4 Set-Up
1)
To assemble the Model S Laboratory Stand, place the upright rod into the base (refer
to assembly instructions in Appendix H). The rack gear and clamp assembly should
face the front of the base. The upright rod is held in place with the jam nut which is
attached from the bottom of the base. Tighten this nut with a suitable wrench
(spanner). Attach leveling feet.
2)
Insert the mounting rod on the back of the DV-I+ Viscometer into the hole on the
clamp assembly. Be sure that the clamp screw, VS-41Y, is loose.