The following rules can be inferred from this graph:
With normal, continuous operation:
·
M12, M13: T fluid + T ambient should remain lower than 110
°
C
·
M14, M15: T fluid + T ambient should remain lower than 100 °C
·
T fluid should lie between 0 and 100
°
C
·
T ambient should lie between 0 and 55
°
C
When cleaning (without electrical power to the instrument):
·
T fluid + T ambient should remain lower than 130
°
C
·
T fluid should lie between 0 and 130
°
C
·
T ambient should lie between 0 and 80
°
C
3.8
Manual controls
The instrument is equipped with two LEDs and a push button switch, which can be used to
monitor the instrument visually and start several functions manually.
3.8.1
LED indications
l
Mode/MOD Operational mode
l
Error/NET
Error/warning indication
The tables below list the different LED indications:
l
Green
Pattern
Time
Indication
off
continuous
Power-off
on
continuous
Normal operation mode
short flash
0.1 sec on,
2 sec off
Initialization mode
blink
0.2 sec on,
0.2 sec off
Special function mode; the instrument is busy performing a special function (e.g.
auto-zero or self-test)
long flash
2 sec on,
0.1 sec off
DeviceNet™
Idle state
Other protocols
n/a