Liquid Dosing Control Manual
9.17.150 rev A – September 2020
Page
26
of
34
Step DDE ProPar Name
Data
Type Execute Value
Remark
Stop Dosing
1
206
33/3
fSetpoint
Uint8 write
0.0
In Capacity Unit.
1
9
1/1
Setpoint
Float write
0
0-32000 = 0-100%
2
130
104/8
Counter Mode
Uint8 write
0
Counter Mode = OFF [Batch mode
= OFF]
4.4 Read parameters
Parameters to be read from the instrument for monitoring or logging purposes.
General
- Read value parameter 129 – Capacity Unit
(Measurement unit, index value)
- Read value parameter 28 – Alarm Info
(Instrument alarm information)
Setpoint
- Read value parameter 206 – Flow Setpoint
(Setpoint in unit)
or
- Read value parameter 9 – Setpoint
(Setpoint 0 - 32000)
Flow
- Read parameter 205 – Flow Measure
(Measure in unit)
or
- Read parameter 8 – Measure
(Measure 0 – 32000)
Temperature
- Read parameter 142 – Temperature
(Temperature in °C)