Operation Manual for Hand-held Refractometer 
Model RBX4582 / RBX5890

 

 
 
Product Layout

 

 
 
 
 
 
 

 
 
 

Specification: 

 

Model: 

RBX4582 

RBX5890  

Measuring Range: 

45 ~ 82% 

58 ~ 90 % 

 

Display Resolution: 

0.5% 

0.2% 

 

 
Temperature Compensation:  Automatic from 10 to 30degree Celsius.  
(ATC model code ending with “A”) 
 

Calibration:

 

 

1. 

Calibration should be made in an ambient environment of 20degree Celsius.  

2. 

Open the 

Light Plate

 & place a tiny drop of calibration fluid in the center of the 

prism. 

 

Model 

 

Calibration fluid 

 

Calibration Point 

RBX4582 

78.8% Sucrose Solution 

78.8% 

RBX5890 

78.8% Sucrose Solution 

78.8% 

 
3. 

Use  the  provided 

Glass  Block

  and  place  the  frosted  side  of  block  on  the 

center of prism where the fluid is.  Ensure the water is spread across the entire 
surface of the 

Glass Block

.  There must not be any bubbles or dry spots on 

the glass block, which is cause by too little liquid use. 

 
 
 
 
 
 
4. 

Wait  for  about  30  seconds,  this  is  to  allow  the  temperature  compensation  to 
take place. 

5. 

Hold the meter in the direction of light source and look into the 

Eyepiece

6. 

Adjust  the 

Focus  Dial

  until  image  is  clear  and  sharp.    Each  person  has  his 

own focusing adjustment.  

7. 

Remove the rubber cap on the 

Calibration knob

, with the use of the provided 

trimming tool and adjust it until the bottom line of the darkest blue section is at 
the brix scale reads precisely the 

Calibration Point above

. Disregard faint or 

white lines 

8. 

Calibration Completed. 

9. 

Rinse the 

Glass Block

 and 

Prism

 with water and wipe them with soft lint-free 

clothes before proceeding with measurement. 

Operation Manual for Hand-held Refractometer 
Model RBX4582 / RBX5890

 

 
 

Product Layout

 

 
 
 
 
 
 

 
 

 
Specification: 

 

Model: 

RBX4582 

RBX5890  

Measuring Range: 

45 ~ 82% 

58 ~ 90 % 

Display Resolution: 

0.5% 

0.2% 

 
Temperature Compensation:  Automatic from 10 to 30degree Celsius.  
(ATC model only) 

 
Calibration:

 

 

1. 

Calibration should be made in an ambient environment of 20degree Celsius.  

2. 

Open the 

Light Plate

 & place a tiny drop of calibration fluid in the center of the 

prism. 

 

Model 

 

Calibration fluid 

 

Calibration Point 

RBX4582 

78.8% Sucrose Solution 

78.8% 

RBX5890 

78.8% Sucrose Solution 

78.8% 

 
3. 

Use  the  provided 

Glass  Block

  and  place  the  frosted  side  of  block  on  the 

center of prism where the fluid is.  Ensure the water is spread across the entire 
surface of the 

Glass Block

.  There must not be any bubbles or dry spots on 

the glass block, which is cause by too little liquid use. 

 
 
 
 
 
 
4. 

Wait  for  about  30  seconds,  this  is  to  allow  the  temperature  compensation  to 
take place. 

5. 

Hold the meter in the direction of light source and look into the 

Eyepiece

6. 

Adjust  the 

Focus  Dial

  until  image  is  clear  and  sharp.    Each  person  has  his 

own focusing adjustment.  

7. 

Remove the rubber cap on the 

Calibration knob

, with the use of the provided 

trimming tool and adjust it until the bottom line of the darkest blue section is at 
the brix scale reads precisely the 

Calibration Point above

. Disregard faint or 

white lines 

8. 

Calibration Completed. 

9. 

Rinse the 

Glass Block

 and 

Prism

 with water and wipe them with soft lint-free 

clothes before proceeding with measurement. 

Eyepiece 

Calibration Knob 

Focus Dial 

Light Plate 

Prism 

Eyepiece 

Calibration Knob 

Focus Dial 

Light Plate 

Prism 

Glass Block

 

Frosted side of  Glass Block

 

 

GLass Block 

Frosted side of  Glass Block 