HD4001 US Issue 01
7
4.8
Your brooder/ICU is supplied with air filter media fitted at the air inlet (see fig 1) and at each end of the
heater enclosure (see picture 9 above). For details of filter replacement see the Servicing section.
4.9
Place the brooder/ICU on a scratch and moisture resistant surface. Worktop height is ideal.
4.10
Connect the mains cable to the inlet. Ensure the connector is pushed fully home in its socket.
4.11
The fan will start and the display will show the current temperature in the cabinet plus a "P" symbol to
show power was interrupted.
4.12
Allow the brooder/ICU to run for at least an hour to stabilise the temperature before introducing animals.
5
Digital Control System
The TLC Eco Series II control system utilises a highly accurate, calibrated sensor for temperature. Be
cautious of low-cost analogue or digital thermometers when comparing them with the brooder/ICU
display readings.
5.1
NORMAL OPERATION – Temperature is continuously displayed.
The asterisk “*” adjacent to the temperature reading shows when the heater power is on. When warming
the asterisk will be continuously on, once warmed up the asterisk will slowly flash as the heater is pulsed
to maintain the correct temperature. When reducing the temperature setting the asterisk may go off, this
is normal.
5.2
POWER LOSS DISPLAY – If power has been interrupted due to a power cut (or when first switching on) a
“P” is shown flashing in the corner of the display. Press the - or + button for 2 or more seconds to clear
the indicator. If the reason for the power loss is not known check the power cable connections are secure.
T*30.0C
T*30.0C
P