11.Do not recharge, use or leave the battery and device in any
high temperature environment such as a location near fire or in
direct sunlight. Doing so may cause the battery to overheat,
ignite or rupture.
12.There is potential hazard from simultaneous connection of a
patient to a high frequency surgical equipment and the
stimulator that may result in burns and possible damage to the
stimulator.
13.Operation in proximity (e.g. 1 m) to a shortwave or
microwave therapy equipment may produce instability in the
stimulator output.
14.When the device is transferred to another patient
,
replace
used pads with new pads.
15.Please avoid the use of the equipment as much as possible
in the following occasions or conditions, as it may result in
incorrect operation: talking on a mobile phone; near the signal
emission base station for broadcasting, television, communica-
tion, radar, navigation, etc.; near active HF SURGICAL
EQUIPMENT and the RF shielded room of an ME SYSTEM for
magnetic resonance imaging.
16.Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of the equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
17.Portable RF communications equipment (including peripher-
als such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the
equipment including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment
could result.
18.The device delivers a current density 0.612 mA/cm2, In order
to mitigate potential adverse events: