1. Product description
Polylux 2 is a stationary light-curing unit with a separate internally mirrored drawer which is used for
processing light-curing materials from bredent (see item 6.1). It is used in the UV range of 350 – 500 nm and
has a power of approx. 20 watts.
1.1 Scope of delivery
Polylux 2 light-curing unit with mirrored drawer, power cord and instructions for use.
2. Indication
Light-curing unit for processing light-curing materials of bredent.
When polymerizing other light-curing materials, please observe the manufacturers‘ recommendations.
Please refer to the respective instructions for use for processing and processing times of the individual
materials.
3. Contraindication
Light-curing veneering materials can not be processed in the Polylux 2 unit!
4. Hazard and safety information
• The voltage of the socket in use must be 220V (European version) or 115V (US version).
• Do not transport the unit together with the polymerization drawer to avoid damage to the lamps.
Liquids or objects should not get into the lamp since they may cause a short circuit.
• Clean the unit only with dry or slightly moist cloths; do not use any solvents. Please unplug the unit prior to
any maintenance work.
• The unit may only be used for the intended use.
• In the event of damage caused by improper operation and/or non-intended use, any liability and warranty
claims will be invalidated.
Prior to adjustment, maintenance, repair or replacement of components, the unit must be disconnected
from all voltage sources if it needs to be opened.
• After maintenance work has been completed, the safety checks (high voltage stability, earthed conductor
test) must be carried out.
• The unit may only be opened by an expert or by a qualifi ed workshop / repair shop.
• When replacing fuses, it must be ensured that only those of the specifi ed type and nominal current rating
are used.
• If it is assumed that the unit can no longer be safely operated, it must be unplugged and ensured that the
unit can not be operated unintentionally. It may be assumed that safe operation is no longer possible
– the unit reveals obvious signs of damage,
– if the unit is no longer working,
– after the unit has been stored under unfavorable conditions for an extended period.
• Use only original spare parts.
• If the unit has been stored in very cold conditions or in high humidity, the unit must be dried at room
temperature or the temperature must be adjusted for one hour before it is operated (without voltage).