BreatheSuite MDI Device User Manual
Copyright 2021 - BreatheSuite Inc (v1.1) 48
Recommended Portable and Mobile RF Communications
Equipment Separation Distances for BreatheSuite MDI
The BreatheSuite MDI Device is intended for use in an
electromagnetic environment in which radiated radiofrequency
(RF) disturbances are controlled. The user and/or installer of the
unit can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile
radio frequency communications equipment (emitters) and the
BreatheSuite MDI Device, according to the maximum output
power of the equipment, as recommended in the table below.
Rated maximum
output power of
transmitter
Minimum separation distance according to the
frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0.01
0.12
0.23
0.1
0.38
0.73
1
1.2
2.3