background image

Quick Reference Guide

Switch On/Off

Connect the Patient Circuit 

    

  

Start/Stop Treatment    

i-Button

Handle the Connectors

Connect the Click-in Battery (Accessory)

This is not a complete guide for the Vivo 45 LS. See your
manual for complete instructions.

1.

Plug the power supply into 
the Vivo.

2.

Press the Start/Stop button 
on the top panel.

If a pre-use test is prompted, 
follow the instructions on the 
screen. 

Always perform a pre-use test when the patient circuit is 
replaced or modified.
 If the pre-use test is not prompted at start up, you can 
start it from the Others section.

1.

Connect the patient circuit to 
the patient air outlet on the 
ventilator.

2.

If having an active exhalation 
valve patient circuit, connect 
the pilot pressure tube at the 
back of the ventilator.

3.

Connect the other end of the tube to the patient inter-
face. 

For leakage circuits, the leak should be at least 12 l/min at 
4 cmH2O to prevent rebreathing of exhaled air.

1.

To start treatment and enter 
operating mode first press 
and hold the Start/Stop but-
ton on the top panel.

2.

Release the Start/Stop but-
ton when the progress bar is 
filled.

3.

To stop treatment and enter standby mode first press 
and hold the Start/Stop button on the front panel.

4.

Release the Start/Stop but-
ton when the progress bar is 
filled.

5.

Press the Audio Pause button 
to stop the treatment.

Use the i-button to get 
explanations about settings and 
alarms.

Make sure that the marking is 
facing up.

Pull the connector sleeve, not the 
cable itself to release the 
connector.

1.

Release the side cover by 
pressing the button under 
the handle.

2.

Open and remove the side 
panel.

3.

Insert the click-in battery. 
Make sure the latch at the 
bottom of the click-in com-
partment is engaged.

4.

Close the side panel. Make 
sure there is a clicking sound 
to secure the side panel.

Reviews: