Usage of Oxygen (Accessory)
Replacement of the Patient Air Inlet Filters
Symbols
Vivo 45 LS
Quick Reference Guide
Doc. 007726 En Y-1
2019-10-29
•
O
2
must be turned off when the Vivo 45 LS is not in
operating mode!
•
It is recommended to monitor the oxygen concen-
trations.
•
Oxygen flow must not exceed 15 l/min.
1.
Mount the oxygen tube to
the connector.
2.
Connect the oxygen tube. A
click sound is heard when the
tube is properly attached.
Disconnect:
1.
Turn off the oxygen source.
2.
Press the release button to
disconnect the oxygen tube.
3.
Remove the tube.
The filters are located in the filter
cassette.
Washable Filter (grey): Replace at
least once a year, wash at least
once a week.
Disposable Filter (white): Replace
at least every 4th week, or more
often in high pollution or pollen-
rich environments.