4.2.5
Breath Rate
The Breath Rate setting defines the minimum number of breaths the Vivo 3 will deliver
as long as no inspiratory trigger effort from the patient is detected. The cycles will be
ventilator-initiated breaths.
Property
Description
Value range
0 BPM to 40 BPM
The available range for Breath Rate might also be limited by
the set values for
Inspiratory Time
and
Target Volume
.
Default value
12 BPM
Setting resolution
1 BPM
Available in modes
PCV+A
PSV
S
S/T
T
4.2.6
Inspiratory Time (Insp. Time)
The Inspiratory Time setting defines the length of each inspiration from start of inspira-
tion to cycling off to expiration.
Property
Description
Value range
0.3 s to 3 s
The available range for Inspiratory Time might also be
limited by the set values for
Breath Rate
.
Default value
1.5 s
Setting resolution
0.1 s
Available in modes
PCV+A
PSV
S/T
T
4.2.7
Minimum Inspiratory Time (Min Insp. Time)
The Minimum Inspiratory Time setting defines a minimum length for each inspiration.
If the Minimum Inspiratory Time is set to Off, the length of the inspiration and/or
maximum inspiratory time is dependent on the set Expiratory Trigger.
Property
Description
Value range
•
Off
•
0.3 s to 3 s
Default value
Off
Treatment Functions and Settings 55
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual