Immunity test
Compliance
Electromagnetic environment
–
Guidance
RF emissions
CISPR 11
Group 1
The Vivo 3 use RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Vivo 3 are suitable for use in all
establishments, including domestic estab-
lishments and those directly connected to
the public low-voltage power supply net-
work that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage fluctua-
tions/flicker
emission
IEC 61000-3-3
Complies
9.7.3
Recommended separation distances between portable and mobile RF
communications equipment and the Vivo 3
The Vivo 3 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Vivo 3 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Vivo 3 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to the frequency of
transmitter (m)
150 kHz to
80 MHz
d= 0.35*
√
P
m
80 MHz to
800 MHz
d= 0.6*
√
P m
800 MHz to
2.5 GHz
d= 1.2*
√
P m
0.01
0.035
0.06
0.12
0.1
0.11
0.19
0.36
1
0.35
0.60
1.2
10
1.1
1.9
3.6
100
3.5
6.0
12
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Notes
•
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual