Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
2–3
PV 101+/PV 102+ Service Manual
MAINTENANCE SERVICE INSTRUCTIONS
2.2.4
Inspection equipment and tools
Before servicing the PV 101+/PV 102+, make sure you have the following equipment at
hand:
•
Test lung
•
Multimeter
•
Measuring equipment for pressure 0–70 mbar
(Thommen HM 28 digital manometer is recommended, BREAS part no. 001934)
•
Flow meter
•
Screwdrivers TX10, TX25, Phillips 1
•
Box socket wrenches M3, M4
•
Allen key 2, 5x25
For the equipment required when calibrating the device, see section 4.2.
2.2.5
Replacement parts
The following parts should be available when servicing the device:
Description
BREAS part no.
Patient tube
000245
Patient air filter, washable (2 pcs)
000996
Patient air filter, disposable (if used)
001259
Air filter, internal cooling
001121
Blower unit assembly, 15.000 hours
003110