Connecting auxiliary monitoring equipment
For monitoring or downloading data the NIPPY 3+ may be connected to a PC or
Laptop computer. The NIPPY 3+ isolated RS232 port is safe for use with any
domestic PC or laptop computer. However, when assembling a system, the
completed system should comply with EN60601-1 (medical systems). For
example, most computers do not comply with this standard, so it should be sited
at a distance, which makes it impossible to touch the computer and the patient at
the same time.
The RS232 port may also be used to run a remote alarm unit.
1. Fresh air inlet
2. Blower (compressor)
3. Flow Sensor
4. Outlet connector
5. EVC Port (IPPV mode only)
6. Breathing Circuit
7. Exhalation Port
8. Patient Connection Port
9. Exhalation Valve (IPPV mode only)
Ventilator System Pneumatic Diagram