Braun ThermoScan Calibration Manual Download Page 2

guide. 

Battery is low, but thermometer will 
still operate correctly.  

Insert new batteries.  

Battery is too low to take correct 
temperature measurements.  

Insert new batteries.  

Anti-theft function of the device is 
enabled; device is locked. Batteries are 
low.  

Return device to station for 
unlocking.  
Insert new batteries.  

See manufacturer's instruction 
guide.  

No probe cover is attached.  

Attach new, clean probe cover. 

No probe cover is attached and 
batteries are low.  

Attach new, clean probe cover. 

Insert new batteries.  

Position error.  

Clear the display by pushing 
the 

Start

 button once.  

Ensure that the position of the 
probe is correct and remains 
stable.  

Ambient temperature is not within the 
allowed operating range of 50 - 104 °F 
(10 - 40 °C) or is changing too rapidly. 

Allow the thermometer to 
remain in a room where the 
temperature is 50 - 104° F (10 - 
40° C) for 30 minutes.  

  

Temperature taken is not within 
typical human temperature range.  

  

HI will be displayed when temperature 
is higher than 108 °F (42.2 °C).  

LO will be displayed when 
temperature is lower than 68 °F (20 °

Ensure that the probe tips and 
lens are clean and a new, clean 
probe cover is attached. Ensure 
that the thermometer is 
properly inserted into the ear 
canal.  

  

Page 2 of 5

Calibration/verification

9/28/2010

file://C:\Documents and Settings\grahamw1\Local Settings\Temp\~hhD427.htm

Summary of Contents for ThermoScan

Page 1: ...ormal operation after device is turned off None Turn device on Display indicates reading stored in memory None See manufacturer s instruction guide Setting used to change the default temperature scale See manufacturer s instruction guide Calibration mode See manufacturer s instruction guide Setting displays hours remaining before anti theft function of the device is enabled and device is locked No...

Page 2: ...art button once Ensure that the position of the probe is correct and remains stable Ambient temperature is not within the allowed operating range of 50 104 F 10 40 C or is changing too rapidly Allow the thermometer to remain in a room where the temperature is 50 104 F 10 40 C for 30 minutes Temperature taken is not within typical human temperature range HI will be displayed when temperature is hig...

Page 3: ...f Turn thermometer on again If error persists reset the thermometer by removing then replacing the batteries Contact Welch Allyn customer service Part number Description Quantity 01802 110 Tester Calibration 9600 Plus North America EMEA 1 01802 119 Tester Calibration 9600 Plus Latin America Asia Pacific 1 01802 220 Tester Calibration 9600 Plus UK 1 4500 925 USB 2 0 5 pin Mini B cable gold 6 feet 1...

Page 4: ...o progress through the test sequence The left detail pane shows the order of the required tests and tracks the test sequence progress As the Welch Allyn Service Tool application completes each test the pass fail indicator changes to green for pass or red for fail and the application proceeds to the next test To terminate the active test sequence at any time click Abort To terminate all tests click...

Page 5: ...on completes all the tests it displays the pass fail test results After you have finished reviewing the test results click Close Page 5 of 5 Calibration verification 9 28 2010 file C Documents and Settings grahamw1 Local Settings Temp hhD427 htm ...

Reviews: