Braun ThermoScan IRT 3020 CO User Manual Download Page 6

To protect the environment, dispose of 
empty batteries at appropriate collec-
tion sites according to national or local 
regulations.

Calibration

The thermometer is initially calibrated at the 
time of manufacture. If this thermometer is used 
according to the use instructions, periodic re-
adjustment is not required. 

Manufacturing date is given by the LOT number 
located in the battery compartment. The first 
number after LOT represents the last digit of the 
year of manufacture.
The last two digits give the week of manufacture. 
An example: LOT 503 – this product has been 
manufactured in week 03 of the year 2005.

Product specifications

Displayed temperature range: 93.2 °F – 108 °F

(34 °C – 42.2 °C)

Operating ambient 
temperature range:

50 °F – 104 °F
(10 °C – 40 °C) 

Display resolution:  

0.1 °F or °C 

Accuracy for patient 

Maximum Laboratory

temperature range 

Error

96.8 °F – 102.2 °F  
(36 °C – 39 °C): 

± 0.4 °F  (± 0.2 °C) 

outside this range: 

± 0.5 °F  (± 0.3 °C) 

Long term storage ranges 

Temperature:

–4 °F to 122 °F 
(–20 °C to 50 °C) 

Humidity:

 95 % non-

condensing 

Battery life:  

2 years / 1000 
measurements

If device is not used within specified temperature 
and humidity ranges the technical accuracy can 
not be ensured.

This infrared thermometer meets requirements 
established in ASTM Standard E 1965-98 (for the 
thermometer system [thermometer with lens filter]).

 

ASTM laboratory accuracy requirements for the 
thermometer only (not including a lens filter) in 
the display range of 96.8 °F to 102.2 °F (36 °C 
to 39 °C)  for infrared thermometers is ± 0.4 °F 
(± 0.2 °C), whereas for mercury-in-glass and 
electronic thermometers, the requirement per 
ASTM Standards E 667-86 and E 1112-86 is  
± 0.2 °F (± 0.1 °C). 

This appliance conforms to the following 
standards: 
DIN EN 60601-1: 3/96 «Medical electrical 
equipment» – Part 1: General requirements for 
safety

DIN EN 60601-1-2/2001 «Medical electrical 
equipment» – Part 1-2: General requirements for safety 
– collateral standard: Electromagnetic compatibility – 
requirements and tests

DIN EN 12470-5: 2003 «Clinical thermometers» – 
Part 5: Performance of infrared ear thermometers 
(with maximum device)