- 61 -
problem
caused by
what to do?
Meter is calibrated against
whole blood from the manu-
facturer.
To check the accuracy of your
system use
Omnitest® 3
Control
solution or measure
against a laboratory device
which uses plasma.
Blood samples are not taken
at the same time.
Battery is flat or there is a
problem with the meter.
Change batteries and if the
problem persists, contact your
distributor or nearest B. Braun
representative.
Insufficient sample or there
is a problem with test strip
or meter.
Apply sufficient amount of a
fresh sample to a new blood
glucose test strip and re-test.
Conduct a control solution
test or a check strip test.
The code numbers of the
meter and of the test strip
vial do not match.
Different
result in
comparison
with another
meter.
The meter
does not
power on.
Test does not
start after
applying
sample.
Re-test with a new test strip.
If the problem persists, please
contact your authorized repre-
sentative or customer support.
There may be a problem
with the test strip.
The result
is doubtful.
Re-test with a new test strip.
Conduct a control solution
test.
Repeat the measurement and
take the samples at the same
time for comparison.