7-5 Lubrication NL4500
NSK PANA SPRAY Plus or other approved lubricant.
Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the handpiece and
spray for approximately 2-3 seconds. Apply lubricant until it expels from the
handpiece head for at least 2 seconds. (Fig. 8)
Spray Nozzle
Handpiece
PANA SPRAY Plus
Fig. 8
• When applying spray be sure to hold the handpiece firmly to prevent the
handpiece from slipping out of the hand due to the spray pressure.
• Hold the spray can upright.
Caution
Summary of Contents for NL4500
Page 26: ......
Page 27: ......
Page 28: ...Brasseler U S A Dental LLC One Brasseler Blvd Savannah GA 31419 U S A...