20
1
Operation 2016-03-07
Measurement Bar
Meter Numbers
Flash Bar
0.5
Blinks ON and OFF
• Occasionally the meter will make a sudden and large movement as soon as the file is inserted into the root canal, but it will return to normal as the file is ad-
vanced down towards the apex.
• The contrary electrode, file electrode and metal parts of the contra angle could cause an adverse reaction if the patient has an allergy to metals. Ask the pa-
tient about this before using the EndoSync.
• Take care that medicinal solutions such as formalin cresol (FC) or sodium hypochlorite do not get on the contrary electrode or the contra angle. These could
cause an adverse reaction such as inflammation.
• The file electrode cannot be used with the following types of files. Use these files without attaching the file electrode. Files with a shank diameter greater
than 1.2 mm, Files with shanks that do not have a circular cross section, Gates Glidden Drills, Tools with large cutting heads such as largo burrs.
• In some cases such as a blocked root canal, a measurement cannot be made. (For details refer to the section of the EndoSync A.I. manual that covers canals
not suitable for measurement.)
• Accurate measurement is not always possible, especially in cases of abnormal or unusual root canal morphology; always take an X-ray to check the measure-
ment results.
• If the meter does not move when the file is inserted, the unit may be malfunctioning and must not be used.
• Do not use an ultra sonic scaler while the contrary electrode is hooked in the patient’s mouth; noise from the scaler could cause the motor to start running re-
sulting in an accident or injury.
• Absolutely never allow the contrary electrode, the handpiece file electrode or the connections for these to contact an ordinary AC power source such as a
socket; this could result in a very serious and dangerous shock.
* Refer to the user manual for the EndoSync A.I. for information about canal measurement and for warnings and notes about
use.
i. The Measurement Bar shows the location of the file tip. The Flash Bar blinks ON and OFF when the file is inside the ca-
nal.
ii. The 0.5 meter reading shows where the file tip is about 0.5 to 1.0 mm from the anatomical apex.
* The numbers 1, 2, and 3 on the meter do not indicate length in millimeters.
iii. If the file tip goes past the Flash Bar, an alarm will sound and the backlight will blink ON and OFF.
Contrary Electrode
Turn ON the EndoSync and EndoSync A.I.
The backlight for the display will be yellow.
Hook the Contrary Electrode in the corner of the pa-
tient’s mouth.
Meter Display
Operation
iii
ii
i
Summary of Contents for EndoSync TR-CM BSL
Page 42: ......
Page 43: ......
Page 44: ...PUB M8084 EA Brasseler Printed in Japan One Brasseler Boulevard Savannah GA 31419 U S A ...