BRASSELER USA EndoDrive PLUS Operation Manual Download Page 4

3

• Compact design.
• Able to operate for approx. 2 continuous hours at the rated load. (Dependent upon use conditions.)
• A large LCD panel enhances ease of use.
• Programmable up to 9 memory settings.
• The auto reverse function is activated depending on the load. A wide variety of functions such as “auto reverse”, 

“auto stop” and “auto reverse off” are available. Combined with such functions nine different operation programs 
are memorable.

• The rotation of the motor handpiece can be activated by the ON/OFF button of the motor handpiece or the Foot 

Control (optional).

• The motor handpiece can be turned on and off by pressing the ON/OFF button alternatively. In addition, ON by 

keep depressed and OFF by releasing it is possible.

• The motor handpiece softly starts. Since it makes a brief stop before it starts in the reverse rotation, there are no 

vibration and shocks during this change over.

• The built in feedback circuit maintains rotation at a constant speed even when the load on the motor handpiece 

changes.

• Has energy saving feature. The power of the motor handpiece automatically turns off when not being used for 10 

minutes. (Auto power-off function)

• The contra angle heads provided for this product are all autoclavable up to 135°C (excluding the motor 

handpiece).

1. FEATURES

• During rotation, the motor handpiece and the motor handpiece cord may affect computer and LAN cable. Noise 

could be heard during operation near a radio receiver.

• The motor handpiece consumes electricity very slightly even when the power supply is off. In addition, 

fully-charged rechargeable batteries, in general, discharge gradually over time even though it is not used. It is 
recommended to recharge the batteries just before use.

• When the motor handpiece stops automatically by detecting a low battery voltage, turning on the power again 

after a while may not detect the low voltage immediately. This is not a failure, but due to battery characteristics. 
The voltage drop does not coincide with the remaining battery capacity.

• Recharge rechargeable batteries after they deplete as much as possible. Repeating short-time use and 

subsequent recharging may shorten their operating time due to the “memory effect.” Batteries may recover after 
repeating complete discharge and full charge for a few times. (This product is equipped with a “Refresh Mode” 
for this purpose.)

• The used Nickel Metal Hydride Batteries are recyclable, but their disposal may sometimes not be permitted by 

the country. Return them to Brasseler USA.

• When disposing of the control unit and Foot Control (Option), follow the instructions of your local government for 

disposal, as they contain materials which may become industrial waste.

• When discarding the contra angle head and motor handpiece, dispose of them as medical waste.
• Users are responsible for operational control, maintenance and inspection.
• Store the system in the place where the temperature is at -10-60°C (14-140°F), humidity at 10-85% RH, 

atmospheric pressure at 500-1060 hPa, and the system is not subject to air with dust, sulfur, or salinity.

• When trouble is found, contact Brasseler USA.
• This product does not consider patient’s age (except infants), gender, weight or nationality.
• No special training is required for this device.

• EndoDrive PLUS should not be used adjacent to or stacked with other equipment and that if adjacent or 

stacked use is necessary, EndoDrive PLUS should be observed to verify normal operation in the 
configuration in which it will be used.

NOTICE

Summary of Contents for EndoDrive PLUS

Page 1: ...Please Read this operation manual carefully before use and keep for future reference OPERATION MANUAL OM E0202E 001 Motor Handpiece System for Root Canal Enlargement ...

Page 2: ... instructions care and maintenance Keep this Operation Manual for future reference Read these safety cautions thoroughly before use and operate the product properly These indicators are to allow you to use the product safely and prevent danger and harm to you and others These are classified by degree of danger damage and seriousness All indicators concern safety and should be followed Cautions for...

Page 3: ...y result in personal injury due to its unexpected breaking or flying off during rotation Do not use or leave the product in a high temperature environment such as under strong direct sunlight in a car under a blazing sun by a fire or near a stove Check the product before use paying attention to looseness vibration noise and temperature heat generation If any abnormal condition is found even slight...

Page 4: ...not used It is recommended to recharge the batteries just before use When the motor handpiece stops automatically by detecting a low battery voltage turning on the power again after a while may not detect the low voltage immediately This is not a failure but due to battery characteristics The voltage drop does not coincide with the remaining battery capacity Recharge rechargeable batteries after t...

Page 5: ...13M DC7V 0 4A Ø20 L108mm 92g including motor handpiece cord Fig 1 2 3 8 10 9 1 12 4 5 6 7 11 Control Unit Operation Panel See Fig 2 Operation Panel and LCD Panel LCD Panel See Fig 2 Operation Panel and LCD Panel Error codes are displayed Power Cord Connector The AC adapter is connected Foot Control Connector The foot control option is connected 1 2 3 4 5 Motor Handpiece Connector The motor handpie...

Page 6: ...on the rotation speed Operation Panel and LCD Panel Fig 2 LCD Panel This indicates that all displays are lighted TORQUE Key The torque limit values can be changed If the speed is to exceed the upper limit or fall below the lower limit the alarm sounds LCD The present torque limit value is displayed The unit is Ncm When it reaches the maximum value is displayed REVERSE Key REVERSE Key TORQUE Key TO...

Page 7: ... available at 1 1 4 1 10 1 16 1 and 20 1 REVERSE Key The rotational direction of the file can be changed by this key It can also be changed while the file is in motion The alarm sounds during reverse rotation LCD The current rotational direction is displayed GEAR RATIO Key A gear ratio can be selected by this key If the correct gear ratio is selected according to the contra angle head in use an ad...

Page 8: ...d regardless of the ON or OFF position of the power LCD The battery mark will always be displayed even if the unit is powered OFF The mark will be animated when the batteries are being charged or in the refresh mode Refer to P 13 6 3 Battery Refresh AC IN Lamp Green The lamp light will always be displayed when power is supplied by the AC adapter even if the unit is powered OFF AUTO REVERSE If the ...

Page 9: ...nector with the mark of the plug down Fig 5 The Foot Control is optional and is not required to operate the EndoDrive PLUS 4 HOW TO CONNECT EACH PART To remove the plug hold the plug ring and pull it out NOTICE Bar Display LCD The bar display graph which shows the degree of load applied to the motor while the file is in motion Fig 3 Fig 5 Foot Control option Plug Foot Control option Plug Foot Cont...

Page 10: ... especially prone to significantly shorter charging and operating times The temperature of the batteries is measured during charging Proper charging cannot be preformed if the charger is placed in an environment which is subject to sharp temperature change next to window subject to direct sunlight near air outlet or fan heater Place it in a place where temperature change is minimal Charging may no...

Page 11: ...he GEAR RATIO key If the correct gear ratio is selected according to the contra angle head in use an adequate rotation speed and torque limit value can be set 4 Select the rotational direction by pressing the REVERSE key The key changes over to the forward rotation F and reverse rotation R each time it is pressed 5 Select the auto reverse mode by pressing the AUTO REVERSE key The key changes over ...

Page 12: ...ou want it to rotate again re press the ON OFF key or re step on the foot control Optional 5 Operation 1 Start the motor handpiece to pressing the ON OFF button briefly Press the button again to stop the handpiece Alternate operation Keep pressing ON OFF button more than a second the motor handpiece starts while the button is pressed If you release the button it stops Intermittent operation If the...

Page 13: ...for more than one second If the program number is changed by the PROGRAM key without pressing the MEMORIZE key the initially memorized preset values remain Cancel function NOTICE 6 Completion of Medical Treatment When the treatment is completed return the motor handpiece to the handpiece stand and hold down the POWER key for more than one second to turn off the power If 10 minutes pass without ope...

Page 14: ...he time of charging At the same time the CHARGE lamp flashes slowly 5 Leave it as it is The batteries are discharged and recharged automatically 6 Hold down the POWER key for a while when you want to stop this process At this time the power does not turn on Lubricate the contra angle head before Calibration This function does not operate if AC adapter is not used Do not touch or apply a load to th...

Page 15: ...or 2 3 seconds Fig 12 2 Cleaning Motor Handpiece When the motor handpiece becomes dirty wipe it off with a cotton cloth moistened with isopropyl alcohol 7 CLEANING Do not lubricate the motor section of the motor handpiece To clean the motor handpiece do not use any solvent such as benzine and thinner Before mounting the lubricated contra angle head to the motor handpiece wipe off extra oil Stand o...

Page 16: ...13 3 Take out the old batteries and pull out the cord holding it at the connector Fig 14 4 Insert the connector of the cord of new batteries into the connector in accordance with the polarity indication label inside the battery chamber and place the batteries into the chamber with care so as not to catch the cord 9 CHANGING BATTERIES In case of difficult to insert the connector onto the battery th...

Page 17: ...nction of the circuit may be suspected Contact Brasseler USA Malfunction of circuit The motor handpiece is locked at the time of the auto reverse mode Malfunction of circuit High load was continuously applied to the motor handpiece for a relatively long time Malfunction of circuit The motor handpiece is locked at the time of startup The voltage of batteries is too low The life of batteries or not ...

Page 18: ...teries is high An error code is displayed Check the connection Check the connection Charge the batteries or use the AC adapter Insert batteries or use the AC adapter Contact Brasseler USA Check the connection Check the connection Contact Brasseler USA Contact Brasseler USA Insert batteries No problem If the temperature of batteries is less than 0 C 32 F the batteries are not rechargeable Charge th...

Page 19: ...d on while pressing the ON OFF button There is a short circuit inside the ON OFF button The power is turned on while pressing the Foot Control Option There is a short circuit inside the Foot Control Option It is rotated by the ON OFF button It is rotated by the Foot Control Option Clean or replace the contra angle head Contact Brasseler USA Check the ON OFF button Contact Brasseler USA Check the F...

Page 20: ...y electronic equipment The EndoDrive PLUS is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supply network that supplies buildings used for domestic purposes TUV Rhineland of North America is a Nationally Recognized Testing Laboratory NRTL in the United States and is accredited by the Standar...

Page 21: ...lectromagnetic immunity The EndoDrive PLUS is intended for use in the electromagnetic environment specified below The customer or the user of the EndoDrive PLUS should assure that it is used in such an environment NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Page 22: ...prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the EndoDrive PLUS as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equ...

Page 23: ... Specifications may be changed without notice ByYour Side in Dentistry One Brasseler Boulevard Savannah GA 31419 To order call 800 841 4522 or fax 888 610 1937 Visit our website www BrasselerUSA com ...