11. Replacing the Turbine
CAUTION
• Make sure that the O-Ring is removed when removing the Turbine. O-Ring might remain in the Head, when removing the Turbine.
If the O-Ring remains, the new Turbine can not be inserted properly.
• Use only Brasseler USA® genuine turbines. If you use other Turbine, specification may not be satisfied and void the warranty.
• The head cap screw thread is very fine. To prevent damage to the thread always first tighten the handpiece head cap using fingers,
then secure with the head cap wrench.
• Make sure that the O-ring (Fig. 10) is correctly located on the cartridge.
1
) Insert a dummy bur into the chuck.
2
)
Unscrew the Head Cap by turning counter-clockwise
and remove.
3
) Push the dummy bur upwards to remove Turbine.
4
) Clean the interior of the Head casing.
5
)
I
nsert the new Turbine.
6
) Replace the Head Cap and firmly screw into place.
SX-PU03: Order Code P1140200
Fig. 10
Wrench
Turbine
Head Cap
Head
O-Ring
PUSH