Recharging the battery
One full charge of the nickel-metal hydride battery allows 8 hours of non-
stop pipetting. A flashing LED indicates the need for recharging.
Recharging indicator
•
LED off:
battery charge is sufficient
•
LED flashes slowly:
remaining battery charge is sufficient for a
maximum of two hours of pipetting; charg-
ing of battery is recommended
•
LED flashes quickly:
finish pipetting activity and charge battery
Use original manufacturer's AC adapter only! Using a different
AC adapter can damage both the instrument and the battery.
Warning!
38