If the buckle is off during measurement, please re-wrap the wrist cuff and fix it tightly and
measure it again.

Do not repeatedly measured within a short time, it may result in wrist gore, so that the
measurement value is not correct. In order to make the arterial blood pressure returned to its
previous state, relax your wrist for 2 or 3 minutes before the next measurement.

If Irregular Heartbeat(IHB) brought by common arrhythmias is detected in the procedure of
blood pressure measurement, a signal of “

”will be displayed.Under this condition, the

blood pressure monitor can keep function,but the results may not be accurate,it’s suggested
that you consult with your physician for accurate assessment.

Interrupt measurement:

If you need to interrupt the measurement for some reason(for example, you feel not good), at

any time. Press “on/off” button. The machine will automatically reduce the air pressure of the
wrist cuff, and interrupt the measurement. If the switch button can not be used, please remove the
wrist cuff and stop the measurement.

5.3 Answers for Questions

Anomalies

Cause

Correction

No display after

press the start button

Whether the power exhausted

Change a new battery

Whether the polarity position

anti-access

Install the battery in proper

placement

Measurement failure

occurs frequently or
measured blood pres

sure value goes extre

mely high or low.

Wristband does not hold with the

heart position relative level

Measuring with correct posture

Wristbands don’t roll up

Roll your wristbands correctly

Whether speak or move the arm

when Measure

Keep quiet, relax body, re-measure

Whether Clothes rolled

oppression arm

Take off clothes that Oppression arm,

re-measure

Pump work, but the

pressure does not

rise

Whether Wrist strap is poor

connections , there is leakage

Ask for repair

Whether Wristbands damaged due

to long years

Please buy a new one

Wrist Cuff leak in a

short time

Whether the cuff wrap too loose

Please wrap the cuff correctly

Even if you press the ON / OFF button , can not enter the

measurement

Please re-power, measured again

During

measurement

suddenly power off

When not use for a long time, the

temperature variation may make

the battery power exhausted

Change a new battery

Summary of Contents for bpw1

Page 1: ...Wrist Type Blood Pressure Monitor User Manual...

Page 2: ...Methods 9 5 2 Operation Steps of Blood Pressure Measurement 10 5 3 Answers for Questions 11 6 Memory function 12 6 1 Memory Recall 12 6 1 1 Read Average Value 12 6 1 2 Read Memory Record 13 6 2 Memor...

Page 3: ...ssure classification function provides convenience for users to judge whether their blood pressure value is normal or not 2 display units kPa mmHg The product will automatically power off in 1 minute...

Page 4: ...mbly This symbol means BF type equipment Caution Diseases such as diabetes hyperlipidemia high blood pressure may accelerate atherosclerosis even can cause obstacle in peripheral circulation Blood pre...

Page 5: ...wise it can not be measured correctly Don t crash and fall off the machine Don t bend or stretch wrist strap Don t pressure the product before the wrist strap is twined around wrist As the dry battery...

Page 6: ...ld not for neonate This blood pressure monitor applies to daily health care for home use or outward activities you can measure your blood pressure duly in order to make blood pressure health managemen...

Page 7: ...ype measurement 2 Result display SBP DBP PR 3 Unit switch Kpa mmHg switch default unit is mmHg 4 Sets of memory Double sets of memory each set has 99 groups memory recall 5 Clock function set the year...

Page 8: ...Battery Installation 1 Take down the battery cover as the direction of the arrow 2 Install two No 7 batteries please pay attention to the positive and negative electrode 3 Cover the battery Attention...

Page 9: ...Press MEMORY key to switch Press SET key to confirm 3 2 Time Setup We can get known from 3 1 press SET key under user mode and it switches to year setting mode the screen displays 2016 and flashes pre...

Page 10: ...ch to BP Unit mode at the same time The system defaults ON if you didn t choose anything Press MEMORY key to switch Press SET key to confirm Note If the instrument is optional without voice no need to...

Page 11: ...t Methods 1 Sit in a chair as shown in pic 1 don t lean forward or bend your body 2 Before measurement take a deep breath three or four times recover to the normal breath relax your shoulders and arms...

Page 12: ...d KPa please refer to 3 4 part for the specific method of setting Take the mmHg value as an example in the following explanation 1 Press the ON OFF key the machine will automatically return to zero ar...

Page 13: ...the wrist cuff and stop the measurement 5 3 Answers for Questions Anomalies Cause Correction No display after press the start button Whether the power exhausted Change a new battery Whether the polari...

Page 14: ...ause Psychological stress wh ich will let your blood pressure test value higher than value under a relax sta te Other anomalies Please press the start button and check again or change a new battery if...

Page 15: ...symptom turns serious Wrist blood pressure values in these patients with upper measure blood pressure may have a greater difference Even healthy people due to inappropriate test conditions there may b...

Page 16: ...r it will expand p eripheral vascular decreased blood pressure this time the wrist blood pressure will decrea sed bigger than the arm blood pressure 8 Manufacturer s Declaration of the EUT Guidance an...

Page 17: ...dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec N A Mains power quality should be that of a typical commercial or hospital environment If the user of the product r...

Page 18: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbo...

Page 19: ...ansmitters and the product as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output o...

Page 20: ...ure monitor 9 2 Maintenance Methods Do not place the unit in the following places Easy splashing Direct sunlight high temperatures humidity dust full of corrosive gas places Some slant vibrative appul...

Page 21: ...ssure test current 500mA without pump and valve working condition Shutdown current 100uA Power 2 AAA alkaline batteries Size 73 6 60 80 mm N W 115g with wristband without battery Working condition Tem...

Page 22: ...Error message and the cause of error correspondence table Table 2 LCD Error message and the cause of error correspondence table LCD Error message The cause of error E0 1 Due to leaks too fast or pulse...

Page 23: ...rating not following the operator s manual 5 The malfunction caused by natural disasters such as flooding fire 6 The malfunction caused by improper repaired by repaired shop which isn t our authorized...

Reviews: