5 Operation
MN031-443-S0-20110112-EN
TissueSeal / TissueSeal PLUS
19
5. Operation
5.1. Before
use
;
The ligation instrument is assembled (see section 4,
page 18) and prepared (see section 7, page 25).
WARNING
Risk of patient injury!
X
Use only approved BOWA ARC generators with
ligation capability (see section 8, page 40).
X
Use only genuine HF cables.
X
Use only suitable products and accessories as
described in the system overview.
X
Use only intact, sterilized products.
WARNING
Risk of patient injuries from the combustion or
explosion of flammable liquids and gases!
X
Carry out electrosurgical operations only with
insufflation using a non-flammable gas (CO
2
).
X
Avoid contact with flammable gases and liquids, such
skin cleaners, disinfectants and anesthetic gases.