8 Reprocessing
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
33
The following information on the allowed number of
reprocessing cycles should be regarded as a guideline.
The actual number may vary depending on the stress level.
The number of reprocessing cycles of the individual instrument parts:
•
Handle: up to 20 times
•
Jaw: up to 20 times
•
Push rod: up to 20 times
•
Blade rod: up to 20 times
•
Cable: up to 20 times
Reprocessing of the vessel sealing instrument comprises the following
steps:
•
Dismantling
•
Soaking
•
Manual removal of contamination
•
Pretreatment in an ultrasonic bath
•
Machine-based reprocessing in a CDM (cleaning and
disinfection machine)
•
Inspection
•
Packing
•
Autoclaving
•
Storage
8.1.
Dismantling
Disassemble the vessel sealing instrument (see section 7, page 28).
Remove the blade from the vessel sealing instrument and discard it
(see section 7.3, page 31)