3 Functionality
900-100_IFU-V1.0_11587-S0-20141209-EN
ARC 100 Operating Manual
23
3.8.2.
EASY neutral electrode monitoring (EASY monitoring)
The EASY monitoring system measures changes in resistance between
the patient and the high-frequency surgery device both before and
during HF activation. If required, it requests personnel to intervene via
an optical and acoustic alarm. For this purpose, a split neutral electrode
with corresponding contact surfaces and suitable transition resistances
which is attached to the patient according to the manufacturer's
instructions is required. The EASY system does not monitor the currents
in the individual sections of split neutral electrodes.
3.9.
Foot switches
In addition to the manual switch, the foot switch can be used to activate
various operating modes.
The foot switches are connected to socket connector
30
.
The following foot switch can be connected to the HF device:
Article No.
Designation
901-012
Single pedal foot switch