9 Maintenance and repair
36
ARC 100 Operating Manual
900-100_IFU-V1.0_11587-S0-20141209-EN
9.
Maintenance and repair
9.1.
Maintenance
DANGER
Infection hazard
To avoid spreading germs and infections, disinfect the surface of
the device and pack it in addition to the shipping packaging before
it leaves the hospital or surgical practice.
Check the device, the device trolley and the accessories (e.g. foot
switch, cable) after each use for damage or defects. In particular,
ensure that the insulation is intact on all cables.
Do not use any damaged device, damaged device trolley or
damaged accessories.
Replace defective accessories immediately.
Have the safety inspection for the device performed once a year.
Consult the appropriate service manual for additional technical
information.
9.1.1.
Safety inspection
Safety inspections must be performed once a year.
The device and accessories may be inspected only by persons
who have the required training, knowledge or experience and who
can perform the inspection independently.
With regard to the safety inspection, you must comply with the
country-specific rules and regulations.
The inspector documents the inspection results and measured values
corresponding to the printed inspection protocol.
In the case of severe deviations from the values of the attached final
acceptance report, or if the specified maximum values were exceeded:
Send the HF device to the service centre; see Section Technical
service, page 38.