A-6
Bovie Medical Corporation
Recommended separation distances between portable and mobile RF
communications equipment and the IDS-400
The IDS-400 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the IDS-400 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the IDS-400 as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =
[
3.5
]
√
P
V
1
80 MHz to 800 MHz
d =
[
3.5
]
√
P
E
1
800 MHz to 2.5 GHz
d =
[
7
]
√
P
E
1
0.12
0.12
0.23
0.38
0.38
0.73
1.2
1.2
2.3
3.8
3.8
7.3
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration – electromagnetic emissions
The IDS-400 is intended for use in the electromagnetic environment listed below. The customer or the
user of the IDS-400 should assure that is is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF Emissions CISPR 11
Group 2
The IDS-400 must emit
electromagnetic energy in order
to perform its intended function.
Nearby electronic equipment may
be affected.
RF Emissions CISPR 11
Class A
The IDS-400 is suitable for use in
all establishments other than
domestic and those directly
connected to the public
low-voltage power supply network
that supplies buildings used in
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Summary of Contents for ids-400
Page 1: ......
Page 2: ...i User s Guide Bovie IDS 400 USER S GUIDE...
Page 3: ...ii Bovie Medical Corporation...
Page 15: ...1 8 Bovie Medical Corporation...
Page 29: ...3 4 Bovie Medical Corporation...
Page 45: ...7 4 Bovie Medical Corporation...
Page 61: ...A 16 Bovie Medical Corporation...
Page 64: ......