Duty Cycle
Under maximum power settings and rated load conditions (Pure cut, 120 watt @ 500
Ω
load), the generator is
suitable for activation times of 10 seconds on, 30 seconds off for one hour.
NOTICE
The internal temperature of the unit is constantly being monitored. If the temperature rises above
85° an alarm sounds, the system displays an error code, and the system disables output power.
Dimensions and Weight
Operating Parameters
Transport and Storage
Audio Volume
The audio levels stated below are for activation tones (bipolar, cut, and coag) and alarm tones (return electrode and
system alarms) at a distance of one meter. Alarm tones meet the requirements for IEC 60601-2-2.
Activation Tone
26
Bovie / Aaron Medical
Width
........................................26 cm (10.25 in.)
Depth
.........................................30.5 cm (12 in.)
Height
......................................15.2 cm (6 in.)
Weight
.......................................< 6.5 kg (< 14 lbs)
Ambient temperature range
10
°
to 40
°
C (50
°
to 104
°
F)
Relative humidity
30% to 75%, noncondensing
Atmospheric pressure
700 to 1060 millibars
Warm-up time
If transported or stored at temperatures outside the operating temperature range,
allow one hour for the generator to reach room temperature before use.
Ambient temperature range
-34
°
to 65
°
C (-29
°
to 149
°
F)
Relative humidity
10% to 100%, condensing
Atmospheric pressure
500 to 1060 millibars
Volume (adjustable)
45 to 65 dBa
Frequency
Cut: 1 kHz
Blend: 1 kHz
Coagulation: 2 kHz
Fulguration: 2 kHz
Bipolar: 2 kHz
Duration
Continuous while the generator is activated
Summary of Contents for Aaron 1250
Page 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Page 8: ...8 Bovie Aaron Medical...
Page 14: ...14 Bovie Aaron Medical...
Page 24: ...24 Bovie Aaron Medical...
Page 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Page 44: ...44 Bovie Aaron Medical...
Page 50: ...50 Bovie Aaron Medical...
Page 62: ...62 Bovie Aaron Medical...