Accessing Patient Options
Physician Trial Kit
9055930-005 Rev A 5 of 86
2. To increase or decrease the stimulation
level of all programmed therapy areas,
press either
p
or
q
from the Level
screen.
3. To change the stimulation level of just
one area:
a) From the Level screen, press
"
as
many times as necessary to find the
desired area by name or number.
b) Press
p
or
q
to make the change.
4. To activate a stimulation program:
a) Press
P
from the Level screen.
b) From the Program screen press
P
as
many times as necessary to highlight
the desired program.
c) Press
p
to activate the selected
program.
To save the new setting in a program,
return to the Program screen, select the
intended program number, and press
q
.
This will replace the existing program with
the new settings.
Accessing Patient Options
Pulse Width (If provided by physician)
1. Press and hold
"
from the Level screen to
activate the Pulse Width Area 1 screen.
2. Press
"
(normal press) as necessary to
cycle to the area you want to adjust, then
press
p
or
q
to increase or decrease
the Pulse Width.