
13
Boston S
cientific (Master Brand DFU
Template 8.2677in x 1
1.6929in
A4, 9223851
9A), EOP MANUAL, MB, REZUM, EN, 51
230807
-0
1A
Black (K) ∆E ≤5.0
Code
Error Title
Error Cause Text
Error Message Instructions
246
Pre-Treatment Failed
High Temperature (Pre-Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery
Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
255
Treatment Halted
Low Temperature (Treatment)
Retract the needle and remove the Delivery Device from patient. Replace Delivery Device.
260
Treatment Halted
High Water Pressure (Treatment)
Check water line for kinks. Resume treatment. If problem persists, replace Delivery Device.
265
Treatment Halted
Low Water Pressure (Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery
Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
270
Syringe is Empty
Syringe Empty
Retract the needle and remove the Delivery Device from patient. Replace syringe and reprime Delivery Device.
275
Prime Failed
Syringe Water Fill Error
Refill syringe and reprime Delivery Device.
280
Treatment Halted
Elevated Coil Temperature
Partial treatment delivered. Check syringe and water line for bubbles or leaks. If no bubbles or leaks are
found, resume treatment. If problem persists, replace Delivery Device. If bubbles are found, replace syringe
and reprime Delivery Device. If leaks are found, replace Delivery Device.
290
Faulty Delivery Device
High Temperature (Idling)
Replace Delivery Device.
291
Faulty Delivery Device
High Water Pressure (Idling)
Replace Delivery Device.
295
Faulty Delivery Device
Needle Deployment Error
Ensure needle is retracted. Replace Delivery Device.
296
Faulty Delivery Device
Needle Retraction Error
Reattempt needle retraction. If problem persists, retract needle manually and replace Delivery Device.
300
Saline Pump Error
Saline Pump Encoder Error
Ensure Delivery Device saline flush line is correctly inserted into saline pump and pump door is closed.
If problem persists, contact Technical Support.
325
Confirm Bladder Drain
Saline Instilled Limit Exceeded
Saline instilled limit exceeded. Please confirm the physician has drained the bladder.
Informational Error Messages
Table 14: Informational Error Message Table
Code
Error Title
Error Cause Text
Error Message Instructions
250
Prime Paused
Vapor Activation Button Released
Delivery Device button was released before priming completed. Press and hold Vapor Activation Button (blue)
to continue priming.
251
Pre-Treatment Paused
Vapor Activation Button Released
Delivery Device button was released before pre-treatment vapor cycle completed. Press and hold Vapor
Activation Button (blue) to restart pretreatment vapor cycle.
341
Prime Paused
RF Power Tolerance Limit Exceeded
Wait for RF Power Supply reset to complete.
342
Pre-Treatment Paused
RF Power Tolerance Limit Exceeded
Wait for RF Power Supply reset to complete.
343
Treatment Halted
RF Power Tolerance Limit Exceeded
Wait for RF Power Supply reset to complete.
41000
Export Error
USB Drive Not Present or Invalid
Re-insert USB flash drive and try again. If problem persists, replace USB flash drive.
41002
Export Error
USB Drive Export Error
Insert a valid USB flash drive with sufficient available memory.
Note:
The error message dialog boxes associated with errors 250 and 251 close when the user engages the Delivery Device vapor button. The error message dialog boxes associated with errors 341,
342, and 343 close automatically when the RF Power Supply reset is complete.