Precision Spectra™ OMG
Directions for Use
Precision Spectra™ OMG
90870349-03 REV C 3 of 72
en
•
For a two-lead trial, connect the
non-Precision Spectra OR Cable
Assemblies into “1-L” (left) and “2-R”
(right). The lead shows up as contacts
1-4 and 9-12 in the Bionic Navigator
3D software. Medtronic Snap-Lid
Connector Cable (Model # 355031)
or Medtronic Multi-lead Trialing Cable
(Model # 355531)
•
If the trial is using the Snap-Lid
Connector Cable, plug the non-
Precision Spectra connector into the
OMG Connector-M socket labeled
“Single Connector.” If 2 four contact-
leads are placed in the Snap-Lid
Connector Cable, the first and second
leads show up as contacts 1-4
and 5-8, respectively, in the Bionic
Navigator 3D software. The 4-contact
leads should be programmed in
Manual Mode.
•
If 2 eight-contact leads are used, the
leads show up as contacts 1-8 and 9-16
in the Bionic Navigator 3D software.
4. Place the OMG Connector in a secure location
on a flat surface so that patient movement will
not cause the connector to fall.
5. Align and connect the “1-L” and “2-R” ports on
the Precision Spectra OMG and the respective
OMG Connector.
6. Making sure that the Precision Spectra ETS is
OFF, connect the Precision Spectra ETS to the
port labeled “CD” on the Precision Spectra OMG.
CAUTION
Failure to turn off the Precision Spectra ETS
before connecting the Precision Spectra OMG
may result in excess stimulation.
Programming the Precision Spectra
ETS
1. Link the Precision Spectra Clinician
Programmer (CP) to the Precision Spectra
ETS per the standard procedure.
2. Use the Bionic Navigator 3D Software to
program the Precision Spectra ETS.
Refer to the
list of competitor leads for guidance in selecting
the programming mode to use with the lead.
Disconnecting the Precision Spectra
OMG
1. Turn OFF stimulation.
2. Disconnect the Precision Spectra ETS from the
Precision Spectra OMG.
3. Disconnect the Precision Spectra OMG from the
OMG Connector.
4. Disconnect the proximal end(s) of the non-
Precision Spectra OR Cable Assembly(ies) from
the OMG Connector.
CAUTION
Failure to turn off the Precision Spectra ETS
before disconnecting the Precision Spectra
OMG may result in excess stimulation.
Cleaning the Precision Spectra OMG
The Precision Spectra OMG can be cleaned using a
mild detergent applied with a lightly dampened cloth
or tissue. Residue from soapy detergents should be
removed with a cloth lightly dampened with water.
Do not use abrasive cleansers for cleaning.