Precision Novi™ Implantable Pulse Generator
Precision Novi™ Implantable Pulse Generator Directions for Use
90930910-07 REV A 4 of 239
en
Specifications and Technical Data
Parameter
Range
Default
Areas (Channels)
4
—
Amplitude
0 – 25.5 mA
0 mA
Rate
2 – 1200 pps
a
40 pps
Width
20 – 1000 μsec
b
210 μsec
Cycle
1 sec – 90 min, OFF
OFF
Ramp ON
1 – 10 sec
3 sec
Contacts
1 – 16, case: +100% to -100%, OFF
1 – 16, case: OFF
a. Only one Area is available if the rate is >130 pps.
b. Amplitude × Width ≤ 12.7 µC.
Materials
Case
Titanium
Header
Epoxy
Strain Relief
Silicone
Size/Volume
70.9 mm x 49.5 mm x 11.3 mm / 33 cm
3
(including header)
Radiopaque Identification Tag
The IPG contains a radiopaque identification tag “BSC S1140”. The identification tag is visible
using standard x-ray proced
ures.