Boston Scientific FLEXTEND Tips Download Page 3

Evaluate electrical performance of the lead 

Verify the electrical performance of the lead 

before

 attaching the lead to the pulse generator and 

after

 allowing sufficient time 

for the effect of local tissue trauma to subside.  If the measurements do not conform to the recommended values listed in 
Table 1, reposition and reaffix the electrode using the positioning technique discussed previously.  
 

 

   

Table 1. Recommended threshold and sensing measurements

a

 

 

 

Atrial data 

Ventricular data 

Voltage threshold

 b

 1.5 V 

 1.0 V 

Current threshold

 b

 1.5 mA 

 1.5 mA 

Intrinsic amplitude 

 2.0 mV 

 5.0 mV 

Impedance 

450-1800 

 450-1800 

 

Source: FLEXTEND Instructions For Use (355229).      

a

 Values measured approximat y 10 minutes after fixation.  

el

b  

Pulse width setting at 0.5 ms.

 

 

  NOTES: 

  Electrical measurements immediately post-fixation may deviate from recommendations. Values may be 

dependant on patient specific factors such as tissue condition, electrolyte balance and drug interactions. 

  Over-rotation of terminal pin may increase local tissue trauma and cause temporarily high thresholds. 

 

Secure the lead 

After successful lead positioning and fixation, secure the lead to the vein using the suture sleeves provided (Figure 3).  Suture 
sleeve tie-down techniques can vary with the lead insertion technique used.   
 

 

Figure 3. Suture sleeve 

 
  
 
  
 

December 13, 2007

                        

 ©2007 Boston Scientific Corporation or its affiliates. All rights reserved.

 

 Page 3 of 3

 

 

Reviews: