Instructions for use
Performing a test
Light bar
Light modes
On
Cartridge is inside.
Blinking
Insert or remove the cartridge.
Pulsing
New test results are available.
• After
5 minutes
the screen is dimmed to
save energy
.
• After
20 minutes
of inactivity the analyser automatically switches to
standby
.
3
Performing a test
3.1 How to find application-specific instructions for use
The application-specific instructions for use for each test
can be found in paper format in each packaging box.
Before you start to perform a test on the analyser, please
make sure you have
read
and
understood
the
application-
specific instructions for use
for the test you want to
perform.
3.2 Login
If passcode protection is enabled, the analyser requires authentication before a test can be started.
Welcome to Vivalytic.
Who are you?
User 2
User 3
User 1
P r o f e s s i o n a l
A d m i n i s t r a to r
P r o f e s s i o n a l
The welcome screen opens automatically when you turn
on the analyser. A list of available users appears.
You can search for a username by scrolling left or right.
Select your
user account
from the user list.
Enter your
passcode
:
• Professional account:
4 digits
• Administrator account: 8 digits
In case of a forgotten passcode, press
forgot
and follow the instructions displayed on the
screen or go to chapter 6.6 for troubleshooting.