background image

Page  6            P/N:  22066P

              © 2004 Bosch

DS240/DS241 Installation Instructions

7.3  Fine Tune Alignment

• Connect the supplied TC2000 Test Cable to Pins 1, 2 and 3 of

P6, the group of pins located just to the right of the receiver’s
terminal strip (see 

Figure 14).

Figure 14 - TC2000 Connection

-  Temporarily disconnect the Remote Indicator Plate connector, if

installed.

• Connect a 20,000 Ohm/volt (or greater) DC VOM to the black

(–) and red (+) leads of the TC2000.

- Set meter scale so that readings of 0.0 to 5.0 VDC can be

read.

• Check the three LEDs on the receiver.

- If the beam is being received by the receiver, the green LED

should be flashing, and the red and yellow LEDs should be on
steady.

- If the green LED is on steady, the beam is not reaching the

receiver. If this is the case, the preliminary alignment proce-
dure should be attempted again.

• Observe the meter readings and adjust the receiver’s optical

module using the Horizontal and Vertical Fine Tuning adjustment
screws for a maximum reading on the meter. This is the most
critical process of alignment. For the most effective operation
of this system, the peak voltage during fine tune alignment
must be obtained.

NOTE: When performing fine tune alignment, keep arms and hands

away from the front of the receiver and out of the beam path.

• The maximum voltage peaked reading will vary depending on the

distance between the transmitter and receiver. The acceptable
peaked voltage range is 0.50 to 5.00 V. The voltage at the
receiver is greater at shorter distances. It is very important to
peak the voltage to insure a stable trouble free detector.

Hint:

Make a note of the alignment voltage. It may be helpful if
future troubleshooting is necessary.

• After completing the fine tune alignment, remove the test cable,

then replace and secure the receiver's cover.

• Check the status of the receiver’s green LED. It should still be

flashing.

• With the meter still connected to the test cable, re-install the test

cable to P6 through the opening in the cover (red lead towards
center of unit).

• An alarm test may be performed here if desired by connecting

the white and black wires of the TC2000. To reset the unit,
temporarily remove power.

• Press the receiver’s Set-up button located just below the P6/test

cable connection (see 

Figure 15). Press only if the covers are

on both transmitter and receiver.

-  This starts a one to two minute automatic internal setup

process that results in a reference voltage for measuring
beam blockages.

Setup
Button

LEDs

Figure 15 - The Receiver's Set-up Button

-  The red and yellow LEDs should turn off, and the green LED

will turn on steady. After some voltage fluctuations, the meter
should settle at 5.0 VDC.

 NOTE: Be careful not to block the beam or move the units during

this time.

- At the end of the setup period, the receiver’s green LED will

flash, and the voltage will drop to between 3.8 and 4.2 VDC.
This is a reference voltage that later readings can be com-
pared to when determining the need for cleaning.

• If the voltage is not between 3.8 VDC and 4.2 VDC, push the Set-

up button again.

• After set-up is complete, remove the test cable (and reconnect

the Remote Indicator Plate connector if used).

• Replace and secure the access door.

8.0

Other Information

8.1 Fire Alarm Reset

To reset the receiver after a fire alarm has occurred, remove power
from the receiver for at least one second, then re-apply (see also,
Section 8.4 Power Outage on page 7).

8.2 Cleaning

At least once a year, the outside of the covers should be cleaned with
a common window cleaner and a soft, clean cloth. Under normal
conditions, this should not result in a trouble alarm as long as the
beam is not continuously blocked for more than 18 sec. After cleaning,
recheck the reference voltage. If the voltage is less than 3.8 VDC, or
greater than 4.2 VDC, calibrate the units (see 

Section 8.3 Reference

Voltage Calibration).

8.3 Reference Voltage Calibration

The reference voltage of the detectors should be checked at least
once a year (or more often if required by local regulations or authorities
having jurisdiction). Clean covers are necessary for proper
reference voltage readings.

Summary of Contents for DS240

Page 1: ...s 22 F to 130 F 30 C to 54 C For UL Listed installations the range is 32 F to 130 F 0 C to 54 C Mounting The units are designed to be mounted to 4 in 10 cm square or octagonal electrical boxes not supplied Listings and Approvals UL S3019 ULC CS692 MSFM Permit 1943 NY City MEA Acceptance MEA274 93 E CSFM 7260 1062 106 FM Job OX2A9 AY Standard Accessories DIS2000 Remote Test Indicator Plate and Test...

Page 2: ...ose to the beam path can cause the beam to become distorted You can test for beam distortion by monitoring the signal voltage After setup the signal voltage of the detector should read 4 VDC Monitor the voltage and turn on all heating and cooling devices in the area The signal voltage should not fluctuate more than 0 20 VDC If it does the detector should be moved to avoid these disturbances 4 2 Av...

Page 3: ... the top edges of the mounting box Over tightening the mounting screws bows the mounting plate inward at the bottom causing the optics to aim low Irregularities in the flatness of the mounting box edge surfaces or screw tabs will worsen this condition Using only mounting boxes with flat regular surfaces and properly formed screw tabs and tightening the mounting screws to secure the mounting plate ...

Page 4: ...ide a means of checking the condition of the detector from ground level If a DIS2000 or DIS240 Remote Indicator Plate will be used install the Remote Indicator Connector to the receiver now see Figure 7 NOTE The DIS240 remote indicator should be located no further than 100 ft 31 m from the receiver and wired with 18 AWG 1 2 mm or larger wire The DIS2000 may be wired up to 500 ft 152 m from the rec...

Page 5: ...e tune the image to the center of the mirror in line with the front and rear bore sights Replace and secure the transmitter s cover when this alignment procedure is complete Check the transmitter s green LED It should be flashing Replace and secure the transmitter s access door Align the receiver to the image of the transmitter using the above preliminary alignment procedure Check the receiver s g...

Page 6: ...rough the opening in the cover red lead towards center of unit An alarm test may be performed here if desired by connecting the white and black wires of the TC2000 To reset the unit temporarily remove power Press the receiver s Set up button located just below the P6 test cable connection see Figure 15 Press only if the covers are on both transmitter and receiver This starts a one to two minute au...

Page 7: ...by setting the Sensitivity Pot Setting as recommended in Section 7 2 Sensitivity Adjustment on page 5 Since the total obscuration of the beam by smoke increases with distance between the units because more smoke will be within the beam path the detector can be made less sensitive as the distance increases Additionally the sensitivity of the detector should be made less at greater distances because...

Page 8: ...esidential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in strict accordance with the manufacturer s instructions and recommendations may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to...

Page 9: ...DS240 DS241 Installation Instructions 2004 Bosch P N 22066P Page 9 9 0 Troubleshooting ...

Page 10: ...Page 10 P N 22066P 2004 Bosch DS240 DS241 Installation Instructions 10 0 Installing Multiple Units ...

Page 11: ...DS240 DS241 Installation Instructions 2004 Bosch P N 22066P Page 11 11 0 Quick Start Installation Flow Chart ...

Page 12: ...osch 22066P 7 04 130 Perinton Parkway Fairport NY 14450 9199 USA Installation Instructions DS240 DS241 Long Range Beam Smoke Detectors Customer Service 800 289 0096 Technical Support 888 886 6189 Page 12 of 12 ...

Reviews: