3.) Release START button.
The device is now in test mode, and the
current pressure is displayed in the SYS
and DIA fields.
4.) Perform test for deviation of pressure
display and pressure circuit integrity
(observe setting time for cuff – at least
30 seconds) in the usual manner.
5.) Press the START button again to switch
off the device.
C) Safety seal
As a safeguard, the upper and lower
parts of the housing can be joined with
a safety seal. Alternatively, a hole dril-
led in the bottom of the monitor for the
connecting screw for the housing can be
sealed for safety.
m
The device must not undergo service or
maintenance while it is in use.