ED01_REV01-2021 IFU OBWB_EN
9- 12
Suction unit insertion and extraction operations (OB2012 model pictured):
•
After grasping the suction unit with two hands as shown in the first photo, place the bottom of the suction unit on
the skid of the wall bracket.
•
Place the device on the right side on the plastic flap (red) as shown in the picture. For the OB1000 suction unit refer
to the left side of the wall bracket.
•
Place the device in a vertical position and with your hand inserted in the upper handle, place your fingers on the
red button and push it downwards. At this point, press the suction unit into place and release the button, which
must slide upwards.
•
Before removing your hands from the suction unit, make sure that it is properly secured (by pulling slightly forward,
the suction unit must be locked). Make sure that the bracket guide is positioned at the highest upper point.
•
The final position of the suction unit is shown in the photo above right.
•
To remove the suction unit from the bracket, put your hand into the handle and use your fingers to press the red
button. With the other hand grasp the device and rotate it slightly downwards to remove it from the bracket.
Never force insertion of the suction unit. Once the device has been secured, check that the battery is
recharged by checking that the indicator light on the front of the suction unit is on (green = end of charge,
yellow = charging in progress.
4.6
Functionality test of the wall bracket
The test operations described in this paragraph allow the user to check the functionality of the support from a
mechanical and electrical point of view. If the test fails or if the state of charge and power supply does not comply,
contact the authorised service centre. This check should be carried out at least once a day at the beginning of your work
shift. Perform these checks monthly or when necessary:
•
Check the operation of the bracket (without suction unit) by repeatedly pressing the lock and unlock button. The
movement must be smooth and there must be no jamming of the mechanism.
•
Always ensure that the fixing screws are fully tightened (using a screwdriver of a suitable diameter for the screw).
•
Ensure that the two red plastic fins are fully inserted into the metal housings provided.
•
Insert the suction unit into the bracket as shown in the previous photos.
•
Visually check that the indicator light on the front of the secretion suction unit is on.
•
Switch on the suction unit and check its operation
WARNINGS
If the test described above is not passed, contact the authorised service department or the
manufacturer immediately. Do not tamper with the mechanical parts, the contact box and/or the
safety spring. Tampering with the wall bracket and/or its mechanical and electrical devices will
seriously compromise the safety and effectiveness of the device in the field.
5
REUSE AND MAINTENANCE
5.1
After docking the device
Always check that the indicator light on the front of the suction unit is on, confirming that the suction unit is powered
and the internal battery is being recharged. Always make sure that the suction unit is properly seated in the wall bracket.
Final position