3 | QUICK MANUAL
QUICK MANUAL
ENDO P® SYSTEM
CONNECTING THE HEATER BAG
Personnel not wearing sterile scrubs, use aseptic technique:
1. Place the three holes of the Heater bag over
the three studs on the left side of the ENDO H®
heater plate Fig. 2a. Place the Heater Bag’s two
holes on the two studs on the right side of the
ENDO H® Fig. 2b.
2. Press firmly to close the ENDO H®.
3. Connect the Endo H® Bag tubes to the Day
Cassette by connecting Fig. 2c to Fig. 1f, then
connect Fig. 2d to Fig. 1e.
4. Check that the cable between ENDO H® and
ENDO P® is correctly connected to the rear of
both devices.
5. Plug the ENDO H® power cord into the earthed
outlet.
6. Turn ON power by using the switch located on the rear of the unit. In a few seconds, the power-
on LED (green) and heating LED (blue) will be lit.
Figure 2
2a
2b
2c
2d
PRIMING THE DAY CASSETTE
1. Make sure that all clamps on the Day
Cassette are open.
2. Make sure the tubing is not kinked or
obstructed by a wheel from for example
the equipment in the operating room.
3. Make sure that the instrument’s inflow and
outflow ports are fully open.
4. The Patient Cassette should not be primed.
5. From the Main Menu press
Prime
and
Start
6. Once the fluid reaches the end of the Day
Cassette tubing press
Standby
to stop the flow.
7. The Day Cassette will not fill completely – this is normal.