CONNECTING THE HEATER BAG

Personnel not wearing sterile scrubs, use aseptic technique:

1.  Place the three holes of the Heater bag over 

the three studs on the left side of the ENDO H® 

heater plate Fig. 2a. Place the Heater Bag’s two 

holes on the two studs on the right side of the 

ENDO H® Fig. 2b.

2.  Press firmly to close the ENDO H®.
3.  Connect the Endo H® Bag tubes to the Day 

Cassette by connecting Fig. 2c to Fig. 1f, then 

connect Fig. 2d to Fig. 1e.

4.  Check that the cable between ENDO H® and 

ENDO P® is correctly connected to the rear of 

both devices.

5.  Plug the ENDO H® power cord into the earthed 

outlet.

6.  Turn ON power by using the switch located on the rear of the unit. In a few seconds, the power-

on LED (green) and heating LED (blue) will be lit.

Figure 2

2a

2b

2c

2d

PRIMING THE DAY CASSETTE    

1.  Make sure that all clamps on the Day 

Cassette are open.

2.  Make sure the tubing is not kinked or 

obstructed by a wheel from for example 

the equipment in the operating room.

3.  Make sure that the instrument’s inflow and 

outflow ports are fully open. 

4.  The Patient Cassette should not be primed.
5.  From the Main Menu press 

Prime

 and 

Start

6.  Once the fluid reaches the end of the Day 

Cassette tubing press 

Standby

 to stop the flow.

7.  The Day Cassette will not fill completely – this is normal.

Summary of Contents for ENDO P SYSTEM

Page 1: ...cm below the patient s median line and approximately 50 cm from the ground When placed on the Bonvisi Tower shown below the ENDO P will be 50 cm from the ground which is ideal Place the ENDO H on the...

Page 2: ...ft Check that the metallic key Fig 1c on the cassette holder is in the horizontal position then slide the cassette into place over the key until it clicks into place and is in contact with the front o...

Page 3: ...o the rear of both devices 5 Plug the ENDO H power cord into the earthed outlet 6 Turn ON power by using the switch located on the rear of the unit In a few seconds the power on LED green and heating...

Page 4: ...NDO P pump Personnel not wearing sterile scrubs 1 If a used check valve remains from a previous procedure this must be removed from the Day Cassette tubing see Fig 3a 2 Connect the Irrigation Tube to...

Page 5: ...n Tube at the Luer lock beneath the check valve Fig 4b Leaving the check valve in place protects the system from contamination when the system is inactive between cases Discard the rest of the Irrigat...

Page 6: ...inuously shown on the device display If not using the ENDO BP the device will use the procedure specific default settings Each surgery type has default settings pre set for flow and pressure If the us...

Page 7: ...1 First single press decreases flow by 30 2 Second single press decrease flow by further 15 1 First double press increases flow by 30 for 2 minutes 2 Second double press increases flow by further 15...

Reviews: