background image

1 | QUICK MANUAL

QUICK MANUAL

ENDO P® SYSTEM

POSITIONING OF ENDO P® AND ENDO H®

When placed on an equipment cart the ENDO P® should be at least 10 cm below the patient's median 

line and approximately 50 cm from the ground. When placed on the Bonvisi Tower (shown below) the 

ENDO P® will be 50 cm from the ground which is 

ideal.

Place the ENDO H® on the lowest shelf.
Place the ENDO P® on the shelf above the ENDO H®.

BONVISI TOWER

Floor

Patient's median line

Distance 

50 cm

Minimum

distance 

10 cm

Summary of Contents for ENDO P SYSTEM

Page 1: ...cm below the patient s median line and approximately 50 cm from the ground When placed on the Bonvisi Tower shown below the ENDO P will be 50 cm from the ground which is ideal Place the ENDO H on the...

Page 2: ...ft Check that the metallic key Fig 1c on the cassette holder is in the horizontal position then slide the cassette into place over the key until it clicks into place and is in contact with the front o...

Page 3: ...o the rear of both devices 5 Plug the ENDO H power cord into the earthed outlet 6 Turn ON power by using the switch located on the rear of the unit In a few seconds the power on LED green and heating...

Page 4: ...NDO P pump Personnel not wearing sterile scrubs 1 If a used check valve remains from a previous procedure this must be removed from the Day Cassette tubing see Fig 3a 2 Connect the Irrigation Tube to...

Page 5: ...n Tube at the Luer lock beneath the check valve Fig 4b Leaving the check valve in place protects the system from contamination when the system is inactive between cases Discard the rest of the Irrigat...

Page 6: ...inuously shown on the device display If not using the ENDO BP the device will use the procedure specific default settings Each surgery type has default settings pre set for flow and pressure If the us...

Page 7: ...1 First single press decreases flow by 30 2 Second single press decrease flow by further 15 1 First double press increases flow by 30 for 2 minutes 2 Second double press increases flow by further 15...

Reviews: