The FreVAX™ device is equipped with several sensors to guarantee an optimal vaccination result. 

8.1.

 

Position Sensor 

This sensor allows for continuous control of the correct holding 

position of the device. Once the trigger has been pushed, before 

allowing liquid to be drawn into the device the position sensor will 

check if it is held upright in order to avoid drawing in air. If the device 

is held in the correct angle, which is indicated by the small circle being 

inside the big circle, liquid from the attached bottle will be drawn in. 

Once the device is fully loaded with liquid, the holding position is no 

longer relevant until the trigger is activated again.  

 

Please note: When using the tubing adapter to deliver vaccine, no position sensor reading is needed 

because there is no bottle mounted on top of the device. In this case, the position sensor can be 

disabled (

user menu 

 settings 

 position sensor 

 off

). 

8.2.

 

Air Detection Sensor 

This sensor measures whether fluid or air is drawn into the device. If 

air is drawn in, e.g. by emptying the bottle and not fitting a new one 

in time, a warning “Air detected” is shown. Afterwards, only rinsing is 

possible, indicated by blue flashing LED bars, until the air exits at the 

injection head. This takes a minimum of 3 rinsing cycles, as the liquid 

circuit has a volume of a little less than 6 ml and one rinsing cycle 

moves a volume of 2 ml. After the rinsing process, you will be able to 

vaccinate again (indicated by white flashing LED bars). 

8.3.

 

Nozzle Sensor 

This sensor enables an extra safety precaution when using the injector. Once the FreVAX™ device is 

ready for vaccination (white flashing LED bars), push and hold the trigger. When the device is applied 

to the skin of the piglet, the safety sleeve around the injection head will slide back, releasing the 

vaccine. If either the trigger is not pushed or the safety sleeve around the injection head does not 

slide back, no vaccine will be released. This allows for a high level of user safety. 

Furthermore, the nozzle sensor measures  if the minimum  contact 

time between device and piglet 

for administration of the full dose

 is 

fulfilled. If the injector is released too early, implicating an underdose, 

the device will give an optical warning via red flashing LED bars and a 

written notification on the display. 

9.

 

Vaccination Process 

 

Press the trigger to load the device. The position sensor indicates 

the position of the bottle in the holder of the FreVAX™. 

Summary of Contents for FreVAX 101

Page 1: ...operating and cleaning 4 4 General description and information 5 4 1 Description 5 4 2 Packaging contents 5 4 3 Operating requirements 5 4 4 Target species 5 4 5 Mode of action 5 5 Bottle connection...

Page 2: ...Rinsing 13 10 2 Cleaning of the device 14 11 Maintenance and Repair 14 11 1 Cylinder Service 14 11 2 Maintenance and repair service 16 12 Dismantling 18 13 The battery and the charger 19 14 Troublesho...

Page 3: ...lly and follow the instructions copyright Henke Sass Wolf GmbH 8300053415 10 20 Subject to amendments in technology and design Printed in Germany The FreVAX injector is manufactured by Henke Sass Wolf...

Page 4: ...other people In case of accidental injection of vaccine and or rinsing fluid seek immediate medical advice and show the vaccine information sheet Beware of the spike in the bottle holder It may contai...

Page 5: ...bottle for use with defined cleaning liquids 5 Universal bottle collars 20mm 30mm 6 Tubing adapter 7 Tubing under the device Threaded bottle collar for use with PET bottle 8 Lubricant for cylinder ser...

Page 6: ...Operation of the FreVAX begins with inserting a bottle of vaccine into the bottle holder A and pressing the trigger F to activate the injector This will start the pump and vaccine is automatically dra...

Page 7: ...ector on a solid surface as shown on the picture and push the bottle onto the spike using the pre punctured hole CAUTION Without pre puncturing of the bottle stopper rubber particles may be drawn into...

Page 8: ...and the FreVAX logos on the display Both logos are shown for 1 second If the battery is sufficiently charged The display will switch initially to the User Menu unless the device is in a loaded conditi...

Page 9: ...re the first injection The fifth page shows the Service Counter which is counting from 12 000 to 0 If the value falls below 0 the counter continues counting negatively Press and hold the left display...

Page 10: ...vaccinating them with min time lapse between applying the device to the skin of the piglet and the injection itself Press and hold the left display button to select the alternative setting Delay s whi...

Page 11: ...Afterwards only rinsing is possible indicated by blue flashing LED bars until the air exits at the injection head This takes a minimum of 3 rinsing cycles as the liquid circuit has a volume of a littl...

Page 12: ...press and hold the trigger to load the device for the first injection During loading Load Device is displayed After loading the device is ready to vaccinate which is indicated by a white flashing LED...

Page 13: ...ce from the skin of the piglet to enable the next loading cycle and injection When kept in contact with the piglet s skin after an injection and still reloading the device will show Please release Noz...

Page 14: ...mercially available disinfectant while wearing rubber gloves and turning moving the safety sleeve around the injection head to remove all traces of dirt Clean the grey housing and the battery either b...

Page 15: ...the head facing downwards in order to drain residues of vaccine or rinsing fluid Step 2 Remove old cylinder Unscrew the old cylinder with a 15 mm wrench Hold the device with the head facing downwards...

Page 16: ...ul completion of cylinder service please reset service counter Starting from the initial screen push the right button three times until reaching menu point Service Counter Press button Reset for at le...

Page 17: ...le during loading to increase suction pressure Farm User 3 If both doesn t help repair through manufacturer Henke Sass Wolf Position sensor 1 Optionally use device without position sensor Farm User 2...

Page 18: ...The FreVAX device may only be dismantled by authorized service engineers In case of unauthorized dismantling the device will lose its warranty Do not open the FreVAX device by yourself due to health...

Page 19: ...ecommended that batteries are not recharged before falling below 50 loading capacity Never store empty batteries as this could damage them and make them unusable Never send ship defective batteries to...

Page 20: ...ions LED lights orange Battery is empty and charging LED lights yellow Battery is partly charged and still charging LED lights green Battery is fully charged The battery charger is not waterproof Oper...

Page 21: ...ot of vaccine on the skin of the animal and no puncture visible as mentioned in chapter 9 it is normal that a small residue of vaccine will remain on the piglet s skin Air inside the tubing system Rin...

Page 22: ...tements CE EU declaration AMB2621 FCC ID R7TAMB2621 Certification Number 5136A AMB2621 16 Disposal The FreVAX device batteries accessories and packaging should be sorted for environment friendly recyc...

Page 23: ...23 17 EC declaration...

Reviews: