9
Troubleshooting
Problem
Device
doesn’t power on
when START/STOP
button is pressed.
or
Cu
ff does n
ot
inflate.
(display remains
off)
Cuff does not
inflate.
(display turns on)
Cuff doesn’t form a
loop around arm.
Possible Reason
Batteries
completely drained.
Cuff not plugged in
correctly.
Cuff not assembled.
Batteries inserted
with incorrect
polarity.
Resolution
Replace batteries
with new
Duracell AA.
Make sure the
connector on cuff
tube is plugged all
the way into the
device.
To form a loop:
pass the end of the
cuff furthest away
from the tubing
through the long
metal D-ring.
The smooth side
(without the hook
and loop fasteners)
should be on the
inside of the cuff
loop.
Remove batteries
and reinstall them
making sure the +
and - polarity
symbols on the
batteries match
those inside the bat-
tery compartment.