Suggested Electrode
Placement Chart
The sticky electrode pads need
to be placed in the correct
position for maximum pain relief.
It is preferable to obtain
assistance from either your
birthing partner or medical
adviser when positioning the
pads. Please read the following
instructions carefully:
1.
Ensure unit remains switched
off. Carefully peel the pads from
the plastic backing using your
forefinger and thumb. Do not pull
on the lead wires.
2.
Position the top two pads
(connected to the left socket
CH1) approximately 4cm either
side of the spine and 4-5cm
above the imaginary navel line as
seen in the diagram.
3.
Position the bottom two pads
(connected to the right socket
CH2) approximately 4cm either
side of the spine and 4-5cm
below the imaginary navel line as
seen in the diagram.
Reference: Cluett E. (1994) SRN
RM ADM PGCEA. Analgesia in
Labour: A review of the TENS
method. Professional Care of
Mother and Child. Mar;4 (2):
50-2.
Bra line
Navel line
Panty line
Using Your Obi TENS +
a)
You are now ready to start
using your Obi TENS +. Switch
the unit on by pressing the
on/off button. The Obi TENS+ will
always start up in BURST 1 mode.
b)
Begin by pressing the
CH1+ key. Each time you press
this key, a small ‘half-moon’
shape will appear on the screen
increasing in number and size as
the intensity is increased. You
will begin to feel a pulsating
sensation which will burst in time
with the small arrow flashing on
the screen.
Ensure that you
increase the intensity slowly.
c)
Repeat the previous step with
the CH2+ key. You will now
feel a sensation in both sets of
pads.
d) BURST 1 (use between
contractions):
This is the first
mode that you will use. It feels
like a bursting, tingling sensation
i.e. the pulses are off and on
periodically. This mode will help
to promote those all-important
natural pain-killing chemicals
known as ‘endorphins’ and
‘encephalins’.
e) BOOST 1 (use during
contractions):
When
experiencing a contraction,
simply press the ‘boost’ button
located on the left side below the
screen. The Obi TENS+ will then
switch automatically into the
BOOST 1 mode for that extra
surge of power required to
combat the pain during
contractions. This will feel like a
continuous sensation.
f )
Once the contraction has
passed, simply press the
‘BOOST’ button. The Obi TENS+
will return to the original BURST
1 mode. Continue this cycle
during your early stages of
labour.
Using the Between Contraction
Timer (BCT) timer
When in BURST mode, the unit
automatically measures your BCT.
This appears at the bottom of the
display. When you switch to
BOOST, the last BCT will remain at
the bottom of the display. The
BCT will always refresh when re-
entering the BURST mode from
the BOOST mode and will record
the next BCT.
N.B. The Obi TENS+ starts up in
BURST Mode. The BCT only
appears after one cycle of
pressing the Boost Button on
and off.
Recalling contraction history
To recall a list of your last 5
‘BCTs’ press the CH1- and
CH2- buttons
simultaneously. The 1st symbol
you will see is a single moon
which is the last BCT.
Use CH2+ to reveal up to 5
BCTs . Use CH2- to
review more recent BCTs.
Recalling the contraction history
will not disable functionality of
the unit. To return to the main
screen press the CH1- and
CH2- buttons simultaneously.
N.B. To avoid losing contraction
history, do NOT switch unit off.
Remove the battereis if you wish
to reset the contraction history.
The Mode Button
There are two modes available
on the Obi TENS+, each with a
BURST and BOOST function.
Mode 1 (Burst 1 and Boost 1)
,
which has been described
above, is primarily used for the
earlier stages of labour.
Mode 2 (Burst 2 and Boost 2)
can be used when your labour
advances and your contractions
become more frequent.
Press the Mode button to
switch between modes. You will
see from the User Display, that
the number next to the BURST or
BOOST will change from 1 to 2
and vice-versa.
NB. Both modes are
interchangeable and can be
used according to your
preference, not necessarily for
the stage of labour you are in.
Mode 2 can be operated in the
same way as mode 1.
Useful Information
1. Continue stimulation for as
long as necessary, unless you
experience discomfort.
2. If you stimulate for a number
of hours ensure the pads do
not become dry.
If this occurs, switch off the
unit, disconnect the
electrodes, and apply a small
drop of water to them.
3. The pads can be used several
times. Replace the pads
when they stop sticking to
the skin. Replacement pads
are available from
www.babycaretens.com.
4. Always ensure that the unit is
switched off before you
remove all the leads and
pads. Do not pull the wires.
ALWAYS remove the jack plug
from its socket by holding
the plug between the
forefinger and thumb and
pulling gently.
5. The Obi TENS + can also be
used as a standard TENS unit
for relief of numerous
conditions, including after
pains and post operative
pains. It is worth bearing this
in mind if you have
a caesarean section.
Please seek medical
advice first.
Post-Natal Pad Placement
Diagrams
KEY:
=first pair of electrodes
=second pair of electrodes.
N.B. Electrodes supplied with this unit
are all the same colour - the black and
white squares show suggested
electrode placement only.
Frequently Asked Questions
Q
When shall I start using my
Obi TENS +?
A
Use as early as you can after
the onset of labour to allow
time for your body’s pain-
killing chemicals to rise.
Q
Can I try the Obi TENS +
before going into labour?
A
Yes, place 2 pads attached to
one leadwire on your forearm
to feel the different pulse
sensations following the
instructions on page 6.
Q
Can I combine the Obi TENS
+ with other medication?
A
Yes, you can still use
analgesics such as gas and
air (entinox) or pethidine.
Q
Can the Obi TENS + be used
in hospital?
A
Yes. However, a midwife may
ask you to switch it off
temporarily.
Q
Can I use the Obi TENS +
if I am considering a water
birth?
A
Yes, but not in the water
itself.
Troubleshooting Questions
Q
Why does the stimulation
not appear to be as strong
after you have used the Obi
TENS + for a while?
A
Increase the intensity, you
may have become used to a
lower setting.
Q
Why does the stimulation
feel strong but ineffective?
A
You may need to switch off
the unit and reposition the
electrode pads.
Q
Why does the stimulation
not feel sufficiently strong
even on a high setting?
A
The batteries may need
replacing.
Q
What do I do if there is no
stimulation, intermittent
stimulation, or weak
stimulation felt even with a
new battery?
A
Check electrode contact.
Check leads.
Check batteries.
Call our customer careline
on +44 (0)208 5329 5959.
0120
July 2017 (1)
Babycare TENS
108 George Lane, South Woodford,
London E18 1AD
Tel (+44) (0)20 8532 9595
Fax (+44) (0)20 8532 9551
www.babycaretens.com
e-mail [email protected]
Fuji Dynamics Ltd, Unit C,3/F, Manley Tower,
No828 Cheung Sha Wan Rd, HK
This manual is recyclable
Babycare TENS is part of the
Body Clock group of companies
Guarantee
Your Obi TENS + machine is guaranteed for a period of 1 year against
manufacturer’s defects. Please record purchase details in the space
provided below for your record purposes. The guarantee does not
include leads, electrodes or battery. Applies to purchases only.
Date of purchase:
Invoice number:
2
2
1
1
Display
TN type LCD
Operation Mode
Burst 1
Burst 2
Boost 1
Boost 2
4 treatment modes:
Burst 1, Burst 2, Boost 1, Boost 2
Pulse (Hz)
Frequency
Pulse Width
(µs)
Description
16
150
1 Burst/sec and
8 pulses per burst
32
150
2 Burst/sec and
8 pulses per burst
80
150
Constant
100
150
Constant
- Output Voltage: 0 – 55V zero to peak at 500Ω load
- Output Waveform: Symmetrical biphasic rectangular
Automatic Power Off
5 Minutes
Output Channel
Two Channel
Output Voltage
0 – 55V (zero-to-peak) across a 500Ω load;
adjustable in 14 steps
Output Intensity
0 – 110 mA (max.) into a 500Ω load
Output Waveform
Symmetrical bi-phasic rectangle
Output Jacks
CE Touch-proof jacks
Load Detector
Contact detection between the device and human skin
(This feature is activated from Level 2 and up)
Battery
2 X 1.5V AAA Battery
Battery Low Detect
2.2V ±0.2V
Tolerances
All output parameters are subjected to a ±20% tolerance
unless other specified.
Weight
52g
Size
40 x 125 x 20mm
Period Pain
Lower Back Pain
Shoulder Pain and Frozen Shoulder
Outer and inner aspect of shoulder
Obi TENS + Technical Data