Think about recording the working channel of each monitor; in case of breakage or failure, this
information will be needed to launch the monitor replace procedure (cf. §7.2.1).
6.7.3
Activate a pill
Note: Before pill activation, please check the monitor's operating channel, the date and time
and the number of data to synchronize.
No modification of these four parameters will be
authorized after activation of a capsule.
In order to activate a capsule, please go to the "Capsule" menu and then the submenu
"
Activation pill
" of the monitor (
Picture 18: Capsule activation menu
After validation of the command "Activation pill", dialog boxes will guide you through the
activation process:
First, the message "Plug the activator" appears on the monitor screen; after connecting
the activator and placing it close to the monitor (<1m), press the button OK as soon as
button appears.
Then the message "Place the pill, small part down" appears. The capsule to activate
has then to be placed in the hole of the activator, big part upwards.
Then press the button OK on the monitor.
Finally, the message "Activation in progress ... Push the activator button " appears. You
must then make a short press on the button of the activator.
Once the button of the Activator activated, the green LED located on it will flash; then
leave the capsule in place and wait until you see the message "Capsule activated, ID
number: XX.XX.XX.XX" on the monitor screen.
It is recommended to note this ID
number
.
Therefore, the capsule is activated and associated with the monitor. Press OK to confirm the
announcement and come back to the Menu Capsule.
Monitor assigns to the capsule a number between 1 and 3, the reference for data display. By
default, the assigned number will always be the lowest available between 1 and 3 (available
mean that there are no associated capsule or stored data associated with this number).
Picture 18: Capsule activation menu