![Body Clock Smart TENS Instructions For Use Manual Download Page 11](http://html.mh-extra.com/html/body-clock/smart-tens/smart-tens_instructions-for-use-manual_2782759011.webp)
Technical Data
Display
8 seven segment digits and
around 95 segments
Operation Mode
Four modes:
• Constant
• Burst - 32Hz-2 bursts per second,
8 pulses per burst
• Massage - 100Hz - fixed frequency
modulating on pulse width
• Mixed - 3 second in mode 1,
3 seconds in mode 2, repeat
Pulse (Hz)
Pulse
Display
Frequency
Width (
µs)
Constant
100
20-250
CONSTANT
Burst
32
20-250
BURST
Massage
100
20-250
MASSAGE
Mixed
32 and 100
20-250
MIXED
Timer
Contdown timer has 5 settings
• Continuous
• 15 Minutes
• 30 Minutes
• 45 Minutes
• 60 Minutes
Automatic Power Off
5 Minutes
Output Channel
Dual Channel
Output Voltage
0-60V adjustable in 15 steps
(500Ω load, peak to 0)
Output Intensity
0 - 120mA adjustable in 15 steps
(500Ω load). Intensity level will drop
back to 0 after mode changes
Output Waveform
Asymmetrical bi-phasic rectangular
Output Jacks
CE touch-proof jacks
Load Detector
Contact detection between the
device and human skin
Battery
2 x 1.5V AA Battery (Type LR6)
Battery Low Detect
2.2V±0.2V
11
Smart TENS Instructions.qx6 30/10/06 16:17 Page