![Body Clock Health Care Elle TENS Instructions For Use Manual Download Page 32](http://html.mh-extra.com/html/body-clock-health-care/elle-tens/elle-tens_instructions-for-use-manual_2782762032.webp)
Specifications
Unit Size
115mm x 55mm x 30mm
Unit Weight
75 grams
Output Jacks
CE Touch-proof jacks
Display
Digital LCD - 8 seven segment digits and around
95 segments
Automatic Power Off
Will switch off after 5 minutes if not used
Battery Low Detect
2.2v ± 0.2v
Battery
2 x 1.5v AA Battery (Type LR6)
Operating Environment
Temperature 0-40˚C
Air humidity 30-75%
Air pressure 700 hPa-1060 hPa
Storage and Transport
Environment
Temperature 0-40˚C
Air humidity 30-90%
Air pressure 700 hPa-1060 hPa
General
Programme
Display
Output Frequency / Duration
Burst
Burst 1
Pulse Frequency (16Hz) Pulse Width (150
애
s)
1 burst per second, 8 pulses per burst*
Burst
Burst 2
Pulse Frequency (32Hz) Pulse Width (150
애
s)
2 bursts per second, 8 pulses per burst*
Boost
Boost 1
Pulse Frequency (80Hz) Pulse Width (150
애
s)
continuous*
Boost 2
Boost 2
Pulse Frequency (100Hz) Pulse Width (150
애
s)
continuous*
Output Channel
Dual Channel
Output Waveform
Symmetrical bi-phasic rectangular
Output Voltage
0 - 55v zero to peak adjustable in 15 steps at 500q load
Output Intensity
0 - 110mA adjustable in 15 steps at 500q load Intensity
level will drop back to zero after programme changes
BIRTH mode
Equipment type
Dual Channel Digital TENS
Classification
This unit is fully compliant with EEC Directive 93/42/EEC,
classified as internally powered equipment type BF.
It is intended for continuous operation.
* Default pulse width after mode changed (pulse width not changed when switched between
BURST and BOOST mode) 5 selectable pulse widths in all modes: 50, 100, 150, 200, 250
32
* Pulse width adjustment does not change when switching between Burst and Boost modes.
Default pulse width is 150
애
s. There are 5 adjustable pulse widths selectable in all modes: 50, 100,
150, 200 and 250
애
s.
Summary of Contents for Elle TENS
Page 1: ...Instructions for use TM ...