
Issue A
Nov 02
PAGE
18
XDS Dry Pump
OPERATION
18
4.7
Shut down
Use the procedure below to shut down the pump:
1.
If shutting the pump down prior to a period of
storage, remove any process gases by running
on high flow gas-ballast for at least one hour.
2.
Close any vacuum system isolation valve to
prevent suckback into the vacuum system
(where fitted).
3.
Switch off the pump using the on/off switch.
When the pump is switched off the fan will
continue to run for 1 minute.
4.
Vent the XDS35i by the gas ballast control or
valve on the inlet.
Summary of Contents for A730-01-983
Page 2: ......
Page 6: ...Issue A Nov 02 XDS Dry Pump PAGE iv iv This page intentionally blank ...
Page 10: ...Issue A Nov 02 XDS Dry Pump PAGE 4 4 This page intentionally blank ...
Page 14: ...Issue A Nov 02 XDS Dry Pump PAGE 8 8 This page intentionally blank ...
Page 16: ...Issue A Nov 02 PAGE 10 XDS Dry Pump INSTALLATION 10 Figure 3 Installation drawing ...
Page 20: ...Issue A Nov 02 XDS Dry Pump PAGE 14 14 This page intentionally blank ...
Page 34: ...Issue A Nov 02 XDS Dry Pump PAGE 28 28 This page intentionally blank ...
Page 36: ...Issue A Nov 02 XDS Dry Pump PAGE 30 30 This page intentionally blank ...
Page 40: ...Issue A Nov 02 XDS Dry Pump PAGE 34 34 This page intentionally blank ...
Page 43: ...This page intentionally blank ...